 "schema_version": 1,
 "id": "msrc_wu2023",
 "name": "MSRC",
 "full_name": "Magnetically steerable robotic catheter with integrated magnetic localisation",
 "group": "Lei Zhou and Farshid Alambeigi (both UT Austin in mohanraj2022; Zhou at Univ. Wisconsin-Madison in wu2023)",
 "group_short": "Wisconsin / UT",
 "family": "magnetic",
 "category": "magnetic-navigation",
 "first_year": 2022,
 "latest_year": 2023,
 "versions": "2022 BioRob: kirigami heat-shrink catheter, Helmholtz pair and friction-drive feeder, manual knobs (mohanraj2022_kirigami, full) -> 2023 RA-L: integrated magnetic localisation (wu2023_integrated)",
 "target_anatomy": "peripheral arterial disease (lower-limb phantoms: Y 20 deg, S-shaped, 30 deg branch)",
 "devices": [
  "catheter"
 ],
 "devices_driven": "Cook G11430 catheter with 40 mm flexure-patterned heat-shrink distal end (3.2 mm pre-shrink OD), tip stack of N52 ring magnets 3 mm OD / 1.6 mm ID / 6 mm",
 "n_dof": 2,
 "n_dof_detail": "2 (insertion and planar steering; mohanraj2022 Sec. I, II-B); hardware: Helmholtz pair (220 mm diameter, 120 turns, up to 11 A) steering; friction-drive feeder on stepper",
 "translation_mechanism": "magnetic",
 "translation_detail": "magnetic (steering) + friction-drive feeder (drive wheel with a thick urethane layer, NEMA 17 stepper, spring preload; mohanraj2022 Sec. II-C, III-A)",
 "rotation_mechanism": "magnetic-field",
 "rotation_detail": "field direction",
 "device_range": "N.R.",
 "footprint_mm": "Helmholtz 220 mm diameter, 110 mm spacing",
 "mass_kg": "N.R.",
 "force_sensing": "16 three-axis magnetoresistive sensors (QMC5883L) locate the tip",
 "haptics_leader": "no (human operator controls insertion)",
 "imaging_compat": "fluoroscopy only at bifurcations; digital microscope in bench tests",
 "sterility": "N.R.",
 "highest_validation": "phantom",
 "validation_detail": "three 3D-printed phantoms; catheter navigates under steering plus localisation; trial counts N.R.",
 "key_numbers": "mean position estimate error 0.95 mm after calibration; latency 400 ms mohanraj2022: 25 mT steering field at 13 A (power-supply limit); maximum tip deflection 12 mm; calculated pushing force 8 N (Sec. III-A, IV); one navigation through a 25 deg bifurcation of a leg phantom in 20 s (Sec. III-B, Fig. 9).",
 "sim_model": "mohanraj2022: analytical Helmholtz-field and dipole-torque model, Ansys Maxwell 3D finite elements for field and torque, SolidWorks finite elements for the flexure bending; tip displacement FE vs experiment RMSE 0.575 and 0.328 mm (Sec. II, IV, Figs. 4, 6)",
 "has_sim_model": true,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "B",
  "kind": "photo",
  "source": "wu2023_integrated Fig. 6 centre (Helmholtz coils with 16-sensor array, catheter feed box)",
  "source_url": "https://doi.org/10.1109/lra.2023.3295297",
  "licence": "publisher copyright (IEEE)",
  "credit": "Wu et al., IEEE Robotics and Automation Letters (2023), Fig. 6. publisher copyright (IEEE).",
  "edits": "native extract (4242 px); crop; 7 labels erased and 4 leader lines erased (LaMa), one of which ran over the white coil face to a sensor; sensor-frame inset and axis triad on white background painted white",
  "file": null
 },
 "pictures": [
  {
   "id": "msrc_wu2023_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Catheter navigating a 3D-printed bifurcation phantom with localization on and off",
   "source_key": "wu2023_integrated",
   "source": "Wu et al., IEEE Robotics and Automation Letters (2023), Fig. 10, p. 7",
   "source_url": "https://doi.org/10.1109/lra.2023.3295297",
   "where": "Fig. 10, p. 7",
   "licence": "publisher copyright"
  },
  {
   "id": "msrc_wu2023_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Earlier setup: Helmholtz coils, friction-drive feeder, control box and microscope",
   "source_key": "mohanraj2022_kirigami",
   "source": "Mohanraj et al., 2022 9th IEEE RAS/EMBS International Conference for Biomedical Robotics and Biomechatronics (BioRob) (2022), Fig. 7, p. 5",
   "source_url": "https://doi.org/10.1109/biorob52689.2022.9925350",
   "where": "Fig. 7, p. 5",
   "licence": "publisher copyright"
  },
  {
   "id": "msrc_wu2023_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Digital-microscope snapshots of the catheter steering into a branch",
   "source_key": "wu2023_integrated",
   "source": "Wu et al., IEEE Robotics and Automation Letters (2023), Fig. 9, p. 7",
   "source_url": "https://doi.org/10.1109/lra.2023.3295297",
   "where": "Fig. 9, p. 7",
   "licence": "publisher copyright"
  },
  {
   "id": "msrc_wu2023_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Kirigami catheter passing a 25-degree bifurcation in a leg phantom",
   "source_key": "mohanraj2022_kirigami",
   "source": "Mohanraj et al., 2022 9th IEEE RAS/EMBS International Conference for Biomedical Robotics and Biomechatronics (BioRob) (2022), Fig. 9, p. 6",
   "source_url": "https://doi.org/10.1109/biorob52689.2022.9925350",
   "where": "Fig. 9, p. 6",
   "licence": "publisher copyright"
  },
  {
   "id": "msrc_wu2023_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "CAD of the friction-drive insertion unit with drive wheel and stepper",
   "source_key": "mohanraj2022_kirigami",
   "source": "Mohanraj et al., 2022 9th IEEE RAS/EMBS International Conference for Biomedical Robotics and Biomechatronics (BioRob) (2022), Fig. 5, p. 3",
   "source_url": "https://doi.org/10.1109/biorob52689.2022.9925350",
   "where": "Fig. 5, p. 3",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "wu2023_integrated",
   "doi": "10.1109/lra.2023.3295297",
   "url": null,
   "title": "Integrated Magnetic Location Sensing and Actuation of Steerable Robotic Catheters for Peripheral Arterial Disease Treatment",
   "year": 2023,
   "venue": "IEEE Robotics and Automation Letters"
  },
  {
   "key": "mohanraj2022_kirigami",
   "doi": "10.1109/biorob52689.2022.9925350",
   "url": null,
   "title": "A Kirigami-Based Magnetically Steerable Robotic Catheter for Treatment of Peripheral Artery Disease",
   "year": 2022,
   "venue": "2022 9th IEEE RAS/EMBS International Conference for Biomedical Robotics and Biomechatronics (BioRob)"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (abstract, hardware, results by grep) mohanraj2022_kirigami full (2026-10-14).",
 "notes": "Uniform field, so steering works only in a plane per bifurcation. F4: hardware origin is mohanraj2022 (same coils, catheter, magnets and feeder).",
 "maintainers_own": false,
 "found_by": "actuated-tip"
}