 "schema_version": 1,
 "id": "liberty_microbot2024",
 "name": "LIBERTY",
 "full_name": "LIBERTY Endovascular Robotic System",
 "group": "Microbot Medical",
 "group_short": "Microbot Medical",
 "family": "commercial",
 "category": "commercial",
 "first_year": 2024,
 "latest_year": 2026,
 "versions": "single version; limited market release 2025-2026, full launch from SIR April 2026",
 "target_anatomy": "peripheral arterial/visceral (PAE, UFE, GAE, Y90, peripheral); neurovascular only a stated future target",
 "devices": [
  "guidewire",
  "microcatheter",
  "catheter"
 ],
 "devices_driven": "guide catheter, microcatheter, guidewire (telescopic); standard 0.014-0.018-in guidewires with compatible microcatheters (2-3 F) and guide catheters (4-6 F) (cornelis2026_in p. 3); in the 20-patient study the robot advanced microwire and microcatheter (0.014-0.016-in wires, 2.7 F Progreat or Truselect), guide catheters (typically 5 F) placed by hand (cornelis2026_in p. 3-5, Fig. 2b); tested 0.016-in microwire, 2.4 F microcatheter, 5 F guide (vidal2024_performance)",
 "n_dof": 3,
 "n_dof_detail": "3 (gao2026_legobot via S8); advance, retract, rotate (vidal2024_performance)",
 "translation_mechanism": "friction-roller",
 "translation_detail": "friction-roller (inferred; mechanism not described in vidal2024_performance)",
 "rotation_mechanism": "not-reported",
 "rotation_detail": "N.R.",
 "device_range": "N.R.",
 "footprint_mm": "palm-size drive (vidal2024_performance); 'handheld' system fixed to the angiography table by a mounting arm (cornelis2026_in p. 2-3, Fig. 1b)",
 "mass_kg": "<1 each for drive and wireless hand controller (vidal2024_performance)",
 "force_sensing": "none",
 "haptics_leader": "no: remote controller operable up to 30 ft (9.1 m) away, e.g. from the control room (cornelis2026_in p. 3, Fig. 6b); wireless handheld per vidal2024_performance",
 "imaging_compat": "fluoroscopy with registered DSA overlay (cornelis2026_in Fig. 2a); drive on a mounting arm fixed to the angiography table (Fig. 1b)",
 "sterility": "fully disposable, single-use system: robotic drive, mounting arm and remote controller (cornelis2026_in p. 2, Fig. 1)",
 "highest_validation": "human",
 "validation_detail": "20 pts treated (21 screened), 3 centres (MSKCC, Miami Cardiac & Vascular Institute, Brigham and Women's), 3 interventional radiologists, 3-day follow-up; Y-90 mapping 10, UFE 3, haemangioma 2, TACE 2, sacral, intercostal and pulmonary 1 each; robot removed before embolisation (cornelis2026_in); 2 pigs x 6 operators (vidal2024_performance); simulator benchmark",
 "key_numbers": "navigation to target without conversion 20/20 (95% CI 83.2-100%); robotic navigation 6.3+-7.6 min (median 3, range 1-32) of 120.1+-34.5 min procedure; fluoroscopy median 7.9 min; operator dose median 1.4 uSv (abstract; 1.46 uSv in Table 2) (cornelis2026_in). Wiring on PCI simulator 27.1 s vs manual 36.9 s vs GRX 62.3 s (winsor2024_performance). FDA 510(k) K243789 decided 2025-09-04.",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "A",
  "kind": "product-photo",
  "source": "https://www.microbotmedical.com/wp-content/uploads/2026/04/LIBERTY-Base_image_v2-scaled.webp (RGBA, found via microbotmedical.com WordPress media API; home page https://www.microbotmedical.com/)",
  "source_url": "https://www.microbotmedical.com/wp-content/uploads/2026/04/LIBERTY-Base_image_v2-scaled.webp",
  "licence": "press image (Microbot Medical)",
  "credit": "https://www.microbotmedical.com/wp-content/uploads/2026/04/LIBERTY-Base_image_v2-scaled.webp (RGBA, found via microbotmedical.com WordPress media API; home page https://www.microbotmedical.com/). Licence: press image (Microbot Medical).",
  "edits": "cropped (547,253)-(2095,1499) of 2560x1708, flattened on white; cut = same crop with the publisher alpha (no SAM)",
  "file": null
 },
 "pictures": [
  {
   "id": "liberty_microbot2024_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "LIBERTY drive on its table arm at the patient, operator holding the remote",
   "source_key": "web:microbotmedical.com",
   "source": "microbotmedical.com, (microbotmedical.com home page)",
   "source_url": "https://www.microbotmedical.com/wp-content/uploads/2026/04/Doctor-looking-at-monitor-1240x827.webp",
   "where": "https://www.microbotmedical.com/wp-content/uploads/2026/04/Doctor-looking-at-monitor-1240x827.webp (microbotmedical.com home page)",
   "licence": "press image"
  },
  {
   "id": "liberty_microbot2024_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "Wireless hand-held LIBERTY remote controller, product view",
   "source_key": "web:microbotmedical.com",
   "source": "microbotmedical.com, (microbotmedical.com home page)",
   "source_url": "https://www.microbotmedical.com/wp-content/uploads/2026/04/Control_Image-scaled.jpg",
   "where": "https://www.microbotmedical.com/wp-content/uploads/2026/04/Control_Image-scaled.jpg (microbotmedical.com home page)",
   "licence": "press image"
  },
  {
   "id": "liberty_microbot2024_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Articulated mounting arm with clamp that holds the LIBERTY drive",
   "source_key": "web:microbotmedical.com",
   "source": "microbotmedical.com, (microbotmedical.com home page)",
   "source_url": "https://www.microbotmedical.com/wp-content/uploads/2026/04/LIBERTY-arm_image-scaled.jpg",
   "where": "https://www.microbotmedical.com/wp-content/uploads/2026/04/LIBERTY-arm_image-scaled.jpg (microbotmedical.com home page)",
   "licence": "press image"
  },
  {
   "id": "liberty_microbot2024_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Drive, mounting arm, hand-held remote and clinical set-up drawing",
   "source_key": "vidal2024_performance",
   "source": "Vidal et al., CardioVascular and Interventional Radiology (2024), Fig. 1, p. 3",
   "source_url": "https://doi.org/10.1007/s00270-024-03686-0",
   "where": "Fig. 1, p. 3",
   "licence": "publisher copyright"
  },
  {
   "id": "liberty_microbot2024_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "photo",
   "caption": "Disposable robotic drive and wireless remote controller, product view",
   "source_key": "cornelis2026_in",
   "source": "Cornelis et al., Journal of Vascular and Interventional Radiology (2026), Fig. 1(a), p. 3",
   "source_url": "https://doi.org/10.1016/j.jvir.2026.108591",
   "where": "Fig. 1(a), p. 3",
   "licence": "publisher copyright"
  },
  {
   "id": "liberty_microbot2024_06",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Drive on its mounting arm fixed to the angiography table (arrow)",
   "source_key": "cornelis2026_in",
   "source": "Cornelis et al., Journal of Vascular and Interventional Radiology (2026), Fig. 1(b), p. 3",
   "source_url": "https://doi.org/10.1016/j.jvir.2026.108591",
   "where": "Fig. 1(b), p. 3",
   "licence": "publisher copyright"
  },
  {
   "id": "liberty_microbot2024_07",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "Operator steering with the hand-held remote from the control room",
   "source_key": "cornelis2026_in",
   "source": "Cornelis et al., Journal of Vascular and Interventional Radiology (2026), Fig. 6(b), p. 6",
   "source_url": "https://doi.org/10.1016/j.jvir.2026.108591",
   "where": "Fig. 6(b), p. 6",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "fda2025_k243789",
   "doi": null,
   "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfPMN/pmn.cfm?ID=K243789",
   "title": "510(k) Premarket Notification K243789: LIBERTY Endovascular Robotic System (Microbot Medical, Ltd.)",
   "year": 2025,
   "venue": null
  },
  {
   "key": "cornelis2026_in",
   "doi": "10.1016/j.jvir.2026.108591",
   "url": null,
   "title": "In Vivo Evaluation of a Disposable Endovascular Robotic System for Arterial Peripheral Vascular Interventions: A Multicenter Feasibility Study",
   "year": 2026,
   "venue": "Journal of Vascular and Interventional Radiology"
  },
  {
   "key": "vidal2024_performance",
   "doi": "10.1007/s00270-024-03686-0",
   "url": null,
   "title": "Performance Evaluation of a Miniature and Disposable Endovascular Robotic Device",
   "year": 2024,
   "venue": "CardioVascular and Interventional Radiology"
  },
  {
   "key": "moschovakizeiger2024_safety",
   "doi": "10.1186/s42155-024-00425-x",
   "url": null,
   "title": "Safety and feasibility study of a novel robotic system in an in vivo porcine vascular model",
   "year": 2024,
   "venue": "CVIR Endovascular"
  },
  {
   "key": "winsor2024_performance",
   "doi": "10.1016/j.jscai.2024.102148",
   "url": null,
   "title": "Performance of a Novel Miniaturized Robotic System in Percutaneous Coronary Intervention: A Preclinical Study",
   "year": 2024,
   "venue": "Journal of the Society for Cardiovascular Angiography & Interventions"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (cornelis2026_in); abstract + web (rest)",
 "notes": "Remote ability: Corewell Health teleintervention collaboration did simulated navigation between two sites 5 miles apart (web summary, simulation only). Neuro not yet. Nasdaq/PR: full launch April 2026, territories 4 to 8, goal 12 by end 2026. | X2 2026-11-08: cornelis2026_in read in full; mechanism and DOF count not described there.",
 "maintainers_own": false,
 "found_by": "commercial"
}