 "schema_version": 1,
 "id": "legobot_gao2026",
 "name": "LegoBot",
 "full_name": "LegoBot",
 "group": "NYU BIRO Lab (H. Su)",
 "group_short": "NYU",
 "family": "academic",
 "category": "academic",
 "first_year": 2026,
 "latest_year": 2026,
 "versions": "2026 modular 2-DOF module; 2-module (4-DOF) and 3-module (6-DOF) assemblies",
 "target_anatomy": "neurovascular (diagnostic angiography, aspiration thrombectomy); coronary and peripheral stated as applicable",
 "devices": [
  "guidewire",
  "microcatheter",
  "catheter"
 ],
 "devices_driven": "guidewire 0.014 in; 2.8 Fr/0.028 in microcatheter; 5 Fr angiographic catheter; 6 Fr aspiration catheter; 8 Fr guide catheter (five off-the-shelf devices tested for buckling load)",
 "n_dof": 6,
 "n_dof_detail": "2 per module; 4 (two modules) or 6 (three modules)",
 "translation_mechanism": "friction-roller",
 "translation_detail": "friction-roller",
 "rotation_mechanism": "rotating-clamp",
 "rotation_detail": "clamped-wheel rotation: hollow-shaft direct-drive motor (0.9 N.m peak) turns the clamped instrument; module coupler adds a second torque input via bevel gears; decoupled from translation by command compensation",
 "device_range": "0.33-3 mm (adjustable clamp)",
 "footprint_mm": "345 x 249 x 115 (three modules, Fig. 6a); module 190 mm wide",
 "mass_kg": "N.R.",
 "force_sensing": "motor side only (current, dual 17-bit encoders); no direct instrument sensing (stated limitation)",
 "haptics_leader": "none; Xbox gamepad teleoperation; haptic feedback not yet implemented",
 "imaging_compat": "N.R. (not tested)",
 "sterility": "not validated (stated future work); preloaded replaceable modules; standard Luer lock / Y-connector interfaces",
 "highest_validation": "phantom",
 "validation_detail": "benchtop: 8 Fr catheter slip test; buckling-load test with 5 devices; silicone neurovascular phantom (Type III arch): angiography with 2 modules and aspiration thrombectomy with 3 modules (clot aspirated); one demonstration each, repeat counts N.R.; one timed module exchange 1 min 50 s",
 "key_numbers": "peak insertion force 5.3 N and torque 8.4 mN.m (8 Fr) vs requirement 4.5 N / 8 mN.m; buckling load 0.1 N (0.014 in wire) to 4.3 N (8 Fr); coupler: 5 full microcatheter rotations vs ~180 deg without; 500 Hz command loop; 345x249x115 mm for 3 instruments vs 360x250x250 mm (2 instruments) of Kantu per Table I",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "cleared",
  "grade": "A",
  "kind": "photo",
  "source": "BIRO Lab, New York University (own photograph)",
  "source_url": null,
  "licence": "lab own (BIRO Lab, New York University)",
  "credit": "Photo: BIRO Lab, New York University.",
  "edits": "",
  "file": "pictures/legobot_gao2026.jpg"
 },
 "pictures": [
  {
   "id": "legobot_gao2026_01",
   "status": "cleared",
   "file": "pictures/extra/legobot_gao2026_01.jpg",
   "thumb": "pictures/thumbs/legobot_gao2026_01.jpg",
   "kind": "in-use",
   "caption": "Three-module robot driving devices into a neurovascular phantom, gamepad and GUI",
   "source_key": "gao2026_legobot",
   "source": "Gao et al., IEEE/ASME Transactions on Mechatronics (2026), lab photo IMG_4635.jpg, unannotated original of Fig. 2",
   "source_url": "https://doi.org/10.1109/TMECH.2026.3695773",
   "where": "lab photo IMG_4635.jpg, unannotated original of Fig. 2",
   "licence": "lab own (collaborator material)"
  },
  {
   "id": "legobot_gao2026_02",
   "status": "cleared",
   "file": "pictures/extra/legobot_gao2026_02.jpg",
   "thumb": "pictures/thumbs/legobot_gao2026_02.jpg",
   "kind": "close-up",
   "caption": "Single guide-catheter module with catheter loaded, gear train visible through window",
   "source_key": "gao2026_legobot",
   "source": "Gao et al., IEEE/ASME Transactions on Mechatronics (2026), lab video IMG_4651.MOV still, t = 0.67 s",
   "source_url": "https://doi.org/10.1109/TMECH.2026.3695773",
   "where": "lab video IMG_4651.MOV still, t = 0.67 s",
   "licence": "lab own (collaborator material)"
  },
  {
   "id": "legobot_gao2026_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "photo",
   "caption": "Guide-catheter, microcatheter and guidewire modules chained with triaxial devices loaded",
   "source_key": "gao2026_legobot",
   "source": "Gao et al., IEEE/ASME Transactions on Mechatronics (2026), Fig. 6(b), p. 6",
   "source_url": "https://doi.org/10.1109/TMECH.2026.3695773",
   "where": "Fig. 6(b), p. 6",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "legobot_gao2026_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "photo",
   "caption": "Top view of the stacked three-module 6-DOF prototype with dimensions",
   "source_key": "gao2026_legobot",
   "source": "Gao et al., IEEE/ASME Transactions on Mechatronics (2026), Fig. 6(a), p. 6",
   "source_url": "https://doi.org/10.1109/TMECH.2026.3695773",
   "where": "Fig. 6(a), p. 6",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "legobot_gao2026_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Two-module configuration performing diagnostic angiography in a silicone phantom",
   "source_key": "gao2026_legobot",
   "source": "Gao et al., IEEE/ASME Transactions on Mechatronics (2026), Fig. 9(a), p. 8",
   "source_url": "https://doi.org/10.1109/TMECH.2026.3695773",
   "where": "Fig. 9(a), p. 8",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "legobot_gao2026_06",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Exploded CAD of guidewire and catheter modules: lid, housing, 2-DOF cassette",
   "source_key": "gao2026_legobot",
   "source": "Gao et al., IEEE/ASME Transactions on Mechatronics (2026), Fig. 5(c), p. 5",
   "source_url": "https://doi.org/10.1109/TMECH.2026.3695773",
   "where": "Fig. 5(c), p. 5",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "legobot_gao2026_07",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Cutaway CAD of the adjustable friction-roller clamping mechanism",
   "source_key": "gao2026_legobot",
   "source": "Gao et al., IEEE/ASME Transactions on Mechatronics (2026), Fig. 4, p. 4",
   "source_url": "https://doi.org/10.1109/TMECH.2026.3695773",
   "where": "Fig. 4, p. 4",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "legobot_gao2026_08",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Guide-catheter module on bench with load cell for insertion and torque tests",
   "source_key": "gao2026_legobot",
   "source": "Gao et al., IEEE/ASME Transactions on Mechatronics (2026), Fig. 8(a)-(b), p. 7",
   "source_url": "https://doi.org/10.1109/TMECH.2026.3695773",
   "where": "Fig. 8(a)-(b), p. 7",
   "licence": "publisher copyright (IEEE)"
  }
 ],
 "sources": [
  {
   "key": "gao2026_legobot",
   "doi": "10.1109/TMECH.2026.3695773",
   "url": null,
   "title": "LegoBot: Modular, compact, device-agnostic robot for endovascular catheter and guidewire manipulation in stroke thrombectomy",
   "year": 2026,
   "venue": "IEEE/ASME Transactions on Mechatronics"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full",
 "notes": "Table I (gao2026_legobot) compares with CorPath GRX, LIBERTY, Magellan, R-One, ICL, KAIST, BIT and Kantu. Table I lists Kantu at 360x250x250 mm; the Kantu abstract says 250x350x250 mm.",
 "maintainers_own": true,
 "found_by": "academic"
}