 "schema_version": 1,
 "id": "kuaoc_park2010",
 "name": "KU force-reflecting",
 "full_name": "Korea University force-reflecting robotic platform for cardiac (EP) catheter navigation",
 "group": "Korea University, Korea Artificial Organ Center (Jun Woo Park, Jaesoon Choi, Kyung Sun) with Hui-Nam Pak (Yonsei University Health System) and Jung Chan Lee (Seoul National University) (park2010_development affiliations)",
 "group_short": "Korea U.",
 "family": "academic",
 "category": "academic",
 "first_year": 2010,
 "latest_year": 2011,
 "versions": "2010 platform (park2010_development, Artif. Organs 34(11), full); 2011 forbidden-region haptic virtual fixture on the same master-slave platform, in vitro phantom (park2011_haptic, Artif. Organs 35(11), full)",
 "target_anatomy": "cardiac electrophysiology",
 "devices": [
  "catheter"
 ],
 "devices_driven": "one conventional cardiac ablation catheter loaded with its handle (park2010 Methods); 2011: 7 Fr Blazer II HTD EP catheter (Boston Scientific) (park2011 Methods)",
 "n_dof": 3,
 "n_dof_detail": "3 (forward/backward translation, roll, tip-bending knob) (park2010 Methods, Fig. 1a); master 5 DOF (the 3 plus pitch and yaw for haptic rendering, Fig. 1b)",
 "translation_mechanism": "linear-stage+clamp",
 "translation_detail": "linear-stage+clamp: the whole catheter compartment (handle with roll and knob joints) rides a linear guide and ball screw driven by a Maxon EC-Max30 brushless motor; resolution 0.002 mm (park2010 Methods)",
 "rotation_mechanism": "handle-rotation",
 "rotation_detail": "roll joint (Maxon EC16) turns the mounted catheter and knob joint as one unit (0.037 deg resolution); knob joint: coupling socket on the bending knob driven by an EC16 and a worm drive (0.00185 deg) (park2010 Methods)",
 "device_range": "N.R.",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "load cell (Dacell UMM-1kgf) between the translation joint and the catheter compartment on a sliding bearing measures the axial reaction force, 0.019 N resolution (park2010 Methods, Results); force can also be estimated from motor current (Abstract)",
 "haptics_leader": "custom 5-DOF haptic master with a catheter-handle-shaped housing: knob read by a magnetic encoder (no force feedback), roll with encoder + Maxon RE13 for feedback, translation on a wire-pulley bar with EC-Max13, plus pitch and yaw; RTLinux-Pro hosts linked by Ethernet, 1 kHz force loop (park2010 Methods); 2011: image-based forbidden-region virtual fixture rendered at the master (park2011)",
 "imaging_compat": "N.R. for X-ray; 2011 in vitro tests used a camera (Canon EOS 500D, 1080p, 20 fps, 0.37 mm/pixel) as a fluoroscopy stand-in (park2011 Methods)",
 "sterility": "N.R.",
 "highest_validation": "phantom",
 "validation_detail": "bench: master/slave synchronisation with step, sinusoidal and manual motion, force-sensor check against a push-pull gauge (park2010 Results); phantom: the catheter inserted by teleoperation into a 600 x 400 mm vessel phantom with human-like inner radii while the virtual fixture warned of wall approach (park2011 Methods, Results); animal tests planned (park2010)",
 "key_numbers": "translation max 1000 mm/s and 83,000 mm/s^2, roll max 138 rad/s and 7,610 rad/s^2 (above the 300-700 mm/s and 11-43 rad/s required); force peaks 0.5 and 0.7 N detected, resolution 0.019 N; joint resolutions 0.002 mm / 0.037 deg / 0.00185 deg; 1 kHz synchronisation loop (park2010 Results, Methods)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "B",
  "kind": "render",
  "source": "moon2014_teleoperated Fig. 2 (slave manipulator with a conventional catheter, CAD), ICCAS 2014, doi:10.1109/ICCAS.2014.6988007",
  "source_url": "https://doi.org/10.1109/ICCAS.2014.6988007",
  "licence": "publisher copyright (IEEE/ICROS)",
  "credit": "Moon et al., 2014 14th International Conference on Control, Automation and Systems (ICCAS 2014) (2014), Fig. 2. publisher copyright (IEEE/ICROS).",
  "edits": "native extract (1896 x 920); trimmed 4-10 px of white border; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "kuaoc_park2010_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "Master manipulator: linkage handle on a haptic device (Asan 2014 version)",
   "source_key": "moon2014_teleoperated",
   "source": "Moon et al., 2014 14th International Conference on Control, Automation and Systems (ICCAS 2014) (2014), Fig. 4, p. 2",
   "source_url": "https://doi.org/10.1109/iccas.2014.6988007",
   "where": "Fig. 4, p. 2",
   "licence": "publisher copyright (IEEE/ICROS)"
  },
  {
   "id": "kuaoc_park2010_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "photo",
   "caption": "2010 slave: ball-screw translation, roll and knob joints, catheter handle inset",
   "source_key": "park2010_development",
   "source": "Park et al., Artificial Organs (2010), Fig. 1(a), p. 1036",
   "source_url": "https://doi.org/10.1111/j.1525-1594.2010.01142.x",
   "where": "Fig. 1(a), p. 1036",
   "licence": "publisher copyright (Wiley / ISAO)"
  },
  {
   "id": "kuaoc_park2010_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "2010 haptic master: pitch-yaw base, translation bar, roll and knob handle",
   "source_key": "park2010_development",
   "source": "Park et al., Artificial Organs (2010), Fig. 1(b), p. 1036",
   "source_url": "https://doi.org/10.1111/j.1525-1594.2010.01142.x",
   "where": "Fig. 1(b), p. 1036",
   "licence": "publisher copyright (Wiley / ISAO)"
  }
 ],
 "sources": [
  {
   "key": "park2010_development",
   "doi": "10.1111/j.1525-1594.2010.01142.x",
   "url": null,
   "title": "Development of a Force‐Reflecting Robotic Platform for Cardiac Catheter Navigation",
   "year": 2010,
   "venue": "Artificial Organs"
  },
  {
   "key": "park2011_haptic",
   "doi": "10.1111/j.1525-1594.2011.01373.x",
   "url": null,
   "title": "Haptic Virtual Fixture for Robotic Cardiac Catheter Navigation",
   "year": 2011,
   "venue": "Artificial Organs"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (park2010_development, park2011_haptic;, )",
 "notes": "Jaesoon Choi later led the Asan roller-cartridge robot (A2 asan_choi2021); inference: an early step of the same researcher's line. | 2026-11-08 X3: both papers read in full; highest_validation bench -> phantom (park2011 in vitro vessel phantom). The 2010 slave photograph (Fig. 1a, greyscale, callouts) is now in the gallery; the main picture stays the 2014 Asan CAD (B) whose identity remains inferred",
 "maintainers_own": false,
 "found_by": "gap"
}