 "schema_version": 1,
 "id": "insa_naudin2019",
 "name": "EVAR platform",
 "full_name": "INSA Lyon / Ampère low-cost robotic platform for endovascular aneurysm repair (EVAR): two stacked conveyors pinch the guidewire, an X-Y table rolls it, simulation-driven haptic leader",
 "group": "Laboratoire Ampère (UMR CNRS 5005), Université de Lyon, INSA Lyon (Iris Naudin, Minh Tu Pham, Richard Moreau, Arnaud Lelevé) (naudin2019_evar, affiliation line)",
 "group_short": "INSA Lyon",
 "family": "academic",
 "category": "academic",
 "first_year": 2019,
 "latest_year": 2019,
 "versions": "single version (naudin2019_evar, AIM 2019); its finite-element model of guide insertion comes from the INSA thesis of G. Mouktadiri 2012 and Mouktadiri et al., J. Biomed. Sci. Eng. 2013 (refs 5-6, not read)",
 "target_anatomy": "abdominal aorta and iliac arteries (EVAR of an abdominal aortic aneurysm); patient-specific silicone phantom printed by additive manufacturing, spring-suspended in a water tank (Sec. II, III-A, Fig. 6)",
 "devices": [
  "guidewire",
  "catheter"
 ],
 "devices_driven": "guidewire (\"guide\"); \"it is also possible to drive a catheter by this system instead of a guide wire\"; one device at a time (Sec. III-A)",
 "n_dof": 2,
 "n_dof_detail": "2 device DOF (translation along z, rotation about z) plus a prismatic gap adjustment (y) that sets the conveyor spacing for the device diameter (Sec. III-A, Fig. 3)",
 "translation_mechanism": "belt/track",
 "translation_detail": "belt/track: two conveyors placed one over the other pinch the guidewire, each driven by a DC motor, coupled so they turn in opposite directions (Sec. III-A, IV-A, Figs. 3-4)",
 "rotation_mechanism": "other",
 "rotation_detail": "other: an X-Y table translates one conveyor over the other, rolling the pinched wire about its own axis (cylindrical joint S1-S2; Sec. III-A, Figs. 3, 5)",
 "device_range": "N.R. (conveyor gap adjustable \"to use guide wires with different diameters\", Sec. III-A)",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "none on the robot; the force sent to the leader is the guide-aorta contact force of an ABAQUS simulation, looked up by inserted length (Sec. III-C, IV-B); the authors state it is not the force at the tool (Sec. V)",
 "haptics_leader": "Geomagic Touch (6 DOF, 3 motorised); pressing a stylus button acts as pinching the guide (clutch), stylus translation and rotation are sent to the conveyors and the X-Y table; renders the simulated contact force opposite to the push (Sec. III-A, III-C, Fig. 7)",
 "imaging_compat": "N.R. (no fluoroscopy in the platform; phantom in a transparent water tank)",
 "sterility": "N.R.",
 "highest_validation": "bench",
 "validation_detail": "bench: conveyor translation tracking of the stylus reference trajectory with a proportional speed loop (Simulink/ControlDesk, Fig. 14) and recording of the rotation reference (Sec. IV-A); haptic lookup coupled to the platform (Sec. IV-B). The silicone aorta phantom is built for guide insertion but no insertion result is reported; real-time control was not achieved with MATLAB/Simulink/dSPACE and was re-written in C afterwards (Sec. V). Alekseeva 2025 Table 4 lists it as TRL3 (ex vivo phantom).",
 "key_numbers": "motor speed command 0-10 V, direction 0/5 V (Sec. IV-A); simulated insertion speed 20 mm/s (Sec. IV-B); exit of contact for an arbitrary 90 deg rotation (Sec. V); tracking error shown only as a curve x10 (Fig. 14), no number reported (N.R.)",
 "sim_model": "yes: ABAQUS finite-element model of the guide inserted in a patient aorta and iliac model (elastic, hyperelastic and linear composite materials; Mouktadiri 2013, ref 6) gives contact forces per inserted length, used as the haptic lookup table (naudin2019_evar Sec. III-B, IV-B, Fig. 8)",
 "has_sim_model": true,
 "sim2real": "none-found",
 "sim2real_detail": "none found (the simulation feeds an offline force lookup; no simulation-trained controller)",
 "picture": {
  "status": "not-cleared",
  "grade": "C",
  "kind": "photo",
  "source": "naudin2019_evar Fig. 4 (two conveyors), p. 3",
  "source_url": "https://doi.org/10.1109/AIM.2019.8868919",
  "licence": "publisher copyright (IEEE)",
  "credit": "Naudin et al., 2019 IEEE/ASME International Conference on Advanced Intelligent Mechatronics (AIM) (2019), Fig. 4. publisher copyright (IEEE).",
  "edits": "native extract; no crop; no edit; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "insa_naudin2019_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Top view of the X-Y table that shifts one conveyor to roll the wire",
   "source_key": "naudin2019_evar",
   "source": "Naudin et al., 2019 IEEE/ASME International Conference on Advanced Intelligent Mechatronics (AIM) (2019), Fig. 5, p. 3",
   "source_url": "https://doi.org/10.1109/AIM.2019.8868919",
   "where": "Fig. 5, p. 3",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "insa_naudin2019_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Silicone aorta and iliac phantom suspended in a water tank",
   "source_key": "naudin2019_evar",
   "source": "Naudin et al., 2019 IEEE/ASME International Conference on Advanced Intelligent Mechatronics (AIM) (2019), Fig. 6, p. 3",
   "source_url": "https://doi.org/10.1109/AIM.2019.8868919",
   "where": "Fig. 6, p. 3",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "insa_naudin2019_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "Geomagic Touch haptic device used as the leader, with joint labels",
   "source_key": "naudin2019_evar",
   "source": "Naudin et al., 2019 IEEE/ASME International Conference on Advanced Intelligent Mechatronics (AIM) (2019), Fig. 7, p. 3 (image credited to Geomagic)",
   "source_url": "https://doi.org/10.1109/AIM.2019.8868919",
   "where": "Fig. 7, p. 3 (image credited to Geomagic)",
   "licence": "publisher copyright (IEEE)"
  }
 ],
 "sources": [
  {
   "key": "naudin2019_evar",
   "doi": "10.1109/AIM.2019.8868919",
   "url": null,
   "title": "A Robotic Platform For Endovascular Aneurysm Repair",
   "year": 2019,
   "venue": "2019 IEEE/ASME International Conference on Advanced Intelligent Mechatronics (AIM)"
  },
  {
   "key": "alekseeva2025_design",
   "doi": "10.1002/rob.22480",
   "url": null,
   "title": "The Design of Endovascular Robots for Cardiac Interventions: A Survey",
   "year": 2024,
   "venue": "Journal of Field Robotics"
  },
  {
   "key": "belikov2024_evolution",
   "doi": "10.1016/j.jestch.2024.101789",
   "url": null,
   "title": "Evolution and current state of robotic catheters for endovascular surgery: A comprehensive review",
   "year": 2024,
   "venue": "Engineering Science and Technology, an International Journal"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (naudin2019_evar); review-only (alekseeva2025_design Table 4 row 1; belikov2024_evolution Fig. 6 NA)",
 "notes": "N2 2026-11-08, from review_comparison §1.2 row 23. Built as a low-cost teaching and trajectory-planning platform for EVAR (Abstract, Sec. I). Alekseeva Fig. 12c draws only its rolling-belt principle. Highest validation is bench, not phantom as Alekseeva says: the phantom exists, but no result in it is reported.",
 "maintainers_own": false,
 "found_by": "gap"
}