 "schema_version": 1,
 "id": "hydraulic_gopesh2021",
 "name": "Hydraulic catheter",
 "full_name": "Hydraulic steerable microcatheter tip (hyperelastic, saline-driven chambers)",
 "group": "Friend lab, UC San Diego (Mechanical and Aerospace Engineering) with UCSD Neurosurgery, Radiology and Surgery; concept co-devised with Bernard Yan, Melbourne Brain Centre (gopesh2021_soft p. 1, 12)",
 "group_short": "UC San Diego",
 "family": "tip",
 "category": "actuated-device",
 "first_year": 2021,
 "latest_year": 2021,
 "versions": "Sci. Robotics 2021 (gopesh2021_soft); patent application US20190209811A1 'Hydraulically driven surgical apparatus'; Mooney-Rivlin constants in Gopesh and Friend, Soft Robot. 2020 (ref. 91); medRxiv preprint 2021.06.28.21258767",
 "target_anatomy": "neuro: in vivo external carotid -> ascending pharyngeal -> parotid artery with coil deployment in the parotid artery (not intracranial); PCOM aneurysm coiling in a silicone phantom (gopesh2021_soft p. 6-9)",
 "devices": [
  "guidewire",
  "microcatheter"
 ],
 "devices_driven": "own hydraulic microcatheter: 900 um outer diameter with four 50 um saline channels and a 400 um lumen, 15 mm steerable tip, 160 cm long Pebax shaft (gopesh2021_soft p. 3-5, 9); used through a 5 F guide with a 0.014-in Synchro2 micro-guidewire and AXIUM PRIME coils (p. 10)",
 "n_dof": null,
 "n_dof_detail": "3D tip orientation by pressurising one or more of four channels; no DOF count (gopesh2021_soft p. 1, 3)",
 "translation_mechanism": "tendon/tip-actuation",
 "translation_detail": "tendon/tip-actuation (hydraulic: four saline channels bend the tip; insertion by hand)",
 "rotation_mechanism": "none",
 "rotation_detail": "none reported: bending direction set by which channel is pressurised; shaft rotation N.R.",
 "device_range": "160 cm; 180-degree bend (curvature 400 1/m) at 350 kPa, rupture at 1200 kPa (gopesh2021_soft p. 4)",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "none",
 "haptics_leader": "no: handheld controller connected by microtubing (p. 9); in vivo a 1-ml syringe per channel locked with a one-way stopcock (p. 6); bench: Elveflow OB-1 pressure controller (p. 9)",
 "imaging_compat": "fluoroscopic roadmap (Siemens Artis Zeego); gold radiopaque markers at three positions and barium-filled Pebax (gopesh2021_soft p. 5, 10)",
 "sterility": "N.R.",
 "highest_validation": "animal",
 "validation_detail": "2 Yucatan miniature pigs (47 kg), one per acute terminal session (gopesh2021_soft p. 11): navigation external carotid -> ascending pharyngeal -> parotid artery and coil deployment; patient-derived silicone phantoms (femoral to ICA; ICA to a PCOM aneurysm) with pulsatile flow, trial count N.R.",
 "key_numbers": "180-degree bend at 350 kPa, rupture at 1200 kPa (safety factor about 4); response time < 1 s through 1.6 m of 50 um tubing; length-to-diameter ratio 10,000:1; no endothelial injury or perforation on angiography (gopesh2021_soft p. 3-10)",
 "sim_model": "ANSYS finite-element model of the tip (two-parameter Mooney-Rivlin) to choose channel radius and radial position; the channel count was fixed at four by the authors (gopesh2021_soft p. 4, 10)",
 "has_sim_model": true,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "B",
  "kind": "photo",
  "source": "https://today.ucsd.edu/news_uploads/steerable_catheter_1.png (UCSD news release, today.ucsd.edu/story/bioinspired-steerable-catheter; right part: hydraulic tip photo)",
  "source_url": "https://today.ucsd.edu/news_uploads/steerable_catheter_1.png",
  "licence": "press image (UC San Diego), terms not stated",
  "credit": "https://today.ucsd.edu/news_uploads/steerable_catheter_1.png (UCSD news release, today.ucsd.edu/story/bioinspired-steerable-catheter; right part: hydraulic tip photo). Licence: press image (UC San Diego), terms not stated.",
  "edits": "crop removing the left fluoroscopy inset",
  "file": null
 },
 "pictures": [
  {
   "id": "hydraulic_gopesh2021_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Hyperelastic steerable tip: mold, fabricated tip and cross-section micrographs",
   "source_key": "gopesh2021_soft",
   "source": "Gopesh et al., Science Robotics (2021), Fig. 2(c-d), PMC9809155",
   "source_url": "https://doi.org/10.1126/scirobotics.abf0601",
   "where": "Fig. 2(c-d), PMC9809155",
   "licence": "publisher copyright (NIH author manuscript PMC9809155)"
  },
  {
   "id": "hydraulic_gopesh2021_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Tip fabrication steps from glass-rod mould to the finished four-channel tip",
   "source_key": "gopesh2021_soft",
   "source": "Gopesh et al., Science Robotics (2021), Fig. 7(A-M), p. 8",
   "source_url": "https://doi.org/10.1126/scirobotics.abf0601",
   "where": "Fig. 7(A-M), p. 8",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "gopesh2021_soft",
   "doi": "10.1126/scirobotics.abf0601",
   "url": null,
   "title": "Soft robotic steerable microcatheter for the endovascular treatment of cerebral disorders",
   "year": 2021,
   "venue": "Science Robotics"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (gopesh2021_soft)",
 "notes": "Hand-controlled: no bedside drive in the paper. The earlier '400 kPa' (web) is not in the paper; 350 kPa bends the tip 180 degrees. | X2 2026-11-08: raised to full text.",
 "maintainers_own": false,
 "found_by": "actuated-tip"
}