 "schema_version": 1,
 "id": "hkumrsafe_lee2018",
 "name": "HKU MR-safe robot",
 "full_name": "HKU MR-safe hydraulic robotic manipulator for MRI-guided intracardiac (EP) catheterization",
 "group": "The University of Hong Kong, Department of Mechanical Engineering (Ka-Wai Kwok; Kit-Hang Lee, Ziyang Dong, Xiaomei Wang) with Imperial College London (Su-Lin Lee, Wayne Luk) and CUHK Prince of Wales Hospital (Alex P. W. Lee)",
 "group_short": "HKU",
 "family": "academic",
 "category": "academic",
 "first_year": 2018,
 "latest_year": 2022,
 "versions": "2018 MR-safe hydraulic manipulator (lee2018_mr, full); 2018 model-based (constant-curvature) vs model-free (Jacobian estimated online from EM tip data, no training) teleoperation with a Novint Falcon leader, 10 participants (wang2018_experimental, full); 2019 high-performance continuous hydraulic motor (dong2019_highperformance, title); 2022 upgraded 3-DoF platform with three-cylinder hydraulic actuation units, multicore FBG shape sensing, MR tracking markers and a learning-based piecewise-constant-curvature model; 800 x 140 x 210 mm, insertion 0-340 mm (dong2022_shape, full)",
 "target_anatomy": "cardiac electrophysiology: left atrium, pulmonary vein isolation",
 "devices": [
  "catheter"
 ],
 "devices_driven": "standard clinical EP ablation catheter (about 1.5 m, diameter < 3 mm) 2022: Thermocool Bi-Directional EP catheter (Biosense Webster), 8 Fr, about 2.67 mm (dong2022 Sec. II-A, IV-A).",
 "n_dof": 4,
 "n_dof_detail": "4 in lee2018 (steering/bending, rotation, coarse insertion, fine insertion; lee2018 Table I); 3 (bending, rotation, insertion) in wang2018 Fig. 1 and dong2022 Fig. 6 caption",
 "translation_mechanism": "other",
 "translation_detail": "other: master-slave hydraulic rolling-diaphragm actuators; coarse insertion through a transmission belt, fine insertion stage 2022: rolling-diaphragm three-cylinder actuation units for rotation and insertion (dong2022 Sec. IV-A, Table I).",
 "rotation_mechanism": "other",
 "rotation_detail": "hydraulic rolling-diaphragm piston pair to gear (rotary) unit",
 "device_range": "clinically used EP catheters held in modular 3D-printed holders (lee2018); 2022 ranges: bending -60 to 60 deg at the catheter tip, rotation -360 to 360 deg, insertion 0-340 mm (dong2022 Table I)",
 "footprint_mm": "780 x 105 x 210 (lee2018); 800 x 140 x 210 (2022 platform, dong2022 Table I)",
 "mass_kg": "3.16 (whole slave body)",
 "force_sensing": "none on the catheter; hydraulic force-transmission efficiency estimated 70%",
 "haptics_leader": "Novint Falcon 3D haptic device with a Unity virtual endoscopic view (wang2018 Sec. \"Robot control interface\"); 2022: hydraulic master units driven by motors in the control room, no haptic leader described (dong2022 Sec. IV-A)",
 "imaging_compat": "MR safe: 1.5 T GE SIGNA, SNR loss < 2%, no artefacts; active MR tracking coils on the catheter 2022: tested in a 1.5 T GE SIGNA; 10 m nylon pipelines at 0.1 MPa (dong2022 Sec. V-G, Table I); MR tracking markers about 0.48 mm (Sec. II-C); the tip electrode causes an image artefact (Sec. V-G).",
 "sterility": "disposable insertion guide (lee2018_mr Fig. 3a)",
 "highest_validation": "ex-vivo",
 "validation_detail": "LA phantom ablation task, 127 targets (lee2018_mr Fig. 8); 10 subjects, model-based vs model-free in random order, 121 target spheres around the pulmonary vein ostium in a soft printed LA phantom (wang2018 Sec. \"Simulated ablation task\"); 2022: pulmonary vein isolation on a slice of swine tissue in a silicone LA phantom with pulsatile flow, 12 of 15 targets reached (80.0 %) (dong2022 Sec. V-B to V-F); in the 1.5 T scanner, marker tracking and feedback control with the LA phantom in a water container, no ablation reported in the scanner (dong2022 Sec. V-G)",
 "key_numbers": "ranges: steering +/-45 deg (0.063 deg), rotation +/-360 deg (0.504 deg), coarse 0-200 mm (0.115 mm), fine 0-30 mm (0.016 mm); about 10 m nylon pipelines 6/4 mm, distilled water <= 0.3 MPa; hysteresis 0.88-1.29 mm for 0.05-0.2 MPa preload; delay 45-66 ms at 15 Hz (lee2018_mr Table I, Sec. III); model-free vs model-based (10 subjects): tip-to-target distance 5.61 vs 6.94 mm (19.1 %, p = 0.08), missed targets 33.3 vs 51.9 % (35.8 %, p = 0.03), maximum continuously missed targets 46.2 % fewer (p = 0.01) (wang2018 Table 1); 2022: shape error 2.33 deg per FBG segment; position error 0.63 mm over 21 segments and 1.53 mm at the tip (max 2.27 mm); path following 0.62 mm mean (2.34 mm max) with the learning-based model vs 1.07 mm (4.27 mm) analytical; autonomous targeting 16.9 s per target (dong2022 Sec. V-A, V-B, V-D, V-E)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found (wang2018: the model-free controller updates its Jacobian online from EM data; dong2022: the network is trained on FBG data of the real catheter, Sec. III-B)",
 "picture": {
  "status": "not-cleared",
  "grade": "B",
  "kind": "photo",
  "source": "wang2018_experimental Fig. 1 (whole MR-safe slave robot on the bench), IJCARS 2018, doi:10.1007/s11548-018-1757-z; native embedded raster p. 3",
  "source_url": "https://doi.org/10.1007/s11548-018-1757-z",
  "licence": "publisher copyright",
  "credit": "Wang et al., International Journal of Computer Assisted Radiology and Surgery (2018), Fig. 1. publisher copyright.",
  "edits": "native extract (1417x506); no crop; LaMa erase of 6 boxes on plain background: text tags Bending, Rotation and EP catheter and the EP-catheter pointer (black curtain), the Insertion tag and double arrow (blue drape), and the white rotation arrow beside the rotation unit, which crossed the blue catheter (catheter line restored by the fill, checked at 100 %); the blue bending arrow on the top housing left in place (it lies on the mechanism); not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "hkumrsafe_lee2018_01",
   "status": "cleared",
   "file": "pictures/extra/hkumrsafe_lee2018_01.jpg",
   "thumb": "pictures/thumbs/hkumrsafe_lee2018_01.jpg",
   "kind": "in-use",
   "caption": "Robotic catheter platform at the 1.5T MRI scanner with LA phantom",
   "source_key": "dong2022_shape",
   "source": "Dong et al., IEEE Transactions on Robotics (2022), Fig. 15(a), p. 14",
   "source_url": "https://doi.org/10.1109/tro.2022.3154691",
   "where": "Fig. 15(a), p. 14",
   "licence": "CC BY 4.0"
  },
  {
   "id": "hkumrsafe_lee2018_02",
   "status": "cleared",
   "file": "pictures/extra/hkumrsafe_lee2018_02.jpg",
   "thumb": "pictures/thumbs/hkumrsafe_lee2018_02.jpg",
   "kind": "photo",
   "caption": "MR-safe slave with three-cylinder actuation unit, master part behind the wave guide",
   "source_key": "dong2022_shape",
   "source": "Dong et al., IEEE Transactions on Robotics (2022), Fig. 6, p. 8",
   "source_url": "https://doi.org/10.1109/tro.2022.3154691",
   "where": "Fig. 6, p. 8",
   "licence": "CC BY 4.0"
  },
  {
   "id": "hkumrsafe_lee2018_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Catheter robot and hydraulic tubes at the MRI bore during the compatibility test",
   "source_key": "lee2018_mr",
   "source": "Lee et al., IEEE/ASME Transactions on Mechatronics (2018), Fig. 4(a), p. 5",
   "source_url": "https://doi.org/10.1109/tmech.2018.2801787",
   "where": "Fig. 4(a), p. 5",
   "licence": "publisher copyright"
  },
  {
   "id": "hkumrsafe_lee2018_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "photo",
   "caption": "Robot body, slave actuation units and electric master actuation units",
   "source_key": "lee2018_mr",
   "source": "Lee et al., IEEE/ASME Transactions on Mechatronics (2018), Fig. 1, p. 2",
   "source_url": "https://doi.org/10.1109/tmech.2018.2801787",
   "where": "Fig. 1, p. 2",
   "licence": "publisher copyright"
  },
  {
   "id": "hkumrsafe_lee2018_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "CAD of the slave: insertion guide, catheter holder, actuation units",
   "source_key": "lee2018_mr",
   "source": "Lee et al., IEEE/ASME Transactions on Mechatronics (2018), Fig. 3(a)-(c), p. 4",
   "source_url": "https://doi.org/10.1109/tmech.2018.2801787",
   "where": "Fig. 3(a)-(c), p. 4",
   "licence": "publisher copyright"
  },
  {
   "id": "hkumrsafe_lee2018_06",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "Control interface with endoscopic view and Novint Falcon input",
   "source_key": "wang2018_experimental",
   "source": "Wang et al., International Journal of Computer Assisted Radiology and Surgery (2018), Fig. 4(b), p. 5",
   "source_url": "https://doi.org/10.1007/s11548-018-1757-z",
   "where": "Fig. 4(b), p. 5",
   "licence": "publisher copyright"
  },
  {
   "id": "hkumrsafe_lee2018_07",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Three-cylinder hydraulic actuation unit at the 1.5T MRI scanner",
   "source_key": "dong2019_highperformance",
   "source": "Dong et al., IEEE Robotics and Automation Letters (2019), Fig. 1, p. 2",
   "source_url": "https://doi.org/10.1109/lra.2019.2899189",
   "where": "Fig. 1, p. 2",
   "licence": "publisher copyright"
  },
  {
   "id": "hkumrsafe_lee2018_08",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "3D model of the three-cylinder MR-safe hydraulic motor",
   "source_key": "dong2019_highperformance",
   "source": "Dong et al., IEEE Robotics and Automation Letters (2019), Fig. 2(b), p. 3",
   "source_url": "https://doi.org/10.1109/lra.2019.2899189",
   "where": "Fig. 2(b), p. 3",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "lee2018_mr",
   "doi": "10.1109/tmech.2018.2801787",
   "url": null,
   "title": "MR Safe Robotic Manipulator for MRI-Guided Intracardiac Catheterization",
   "year": 2018,
   "venue": "IEEE/ASME Transactions on Mechatronics"
  },
  {
   "key": "wang2018_experimental",
   "doi": "10.1007/s11548-018-1757-z",
   "url": null,
   "title": "Experimental validation of robot-assisted cardiovascular catheterization: model-based versus model-free control",
   "year": 2018,
   "venue": "International Journal of Computer Assisted Radiology and Surgery"
  },
  {
   "key": "dong2019_highperformance",
   "doi": "10.1109/lra.2019.2899189",
   "url": null,
   "title": "High-Performance Continuous Hydraulic Motor for MR Safe Robotic Teleoperation",
   "year": 2019,
   "venue": "IEEE Robotics and Automation Letters"
  },
  {
   "key": "dong2022_shape",
   "doi": "10.1109/tro.2022.3154691",
   "url": null,
   "title": "Shape Tracking and Feedback Control of Cardiac Catheter Using MRI-Guided Robotic Platform—Validation With Pulmonary Vein Isolation Simulator in MRI",
   "year": 2022,
   "venue": "IEEE Transactions on Robotics"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (lee2018_mr, wang2018_experimental, dong2022_shape); title (dong2019_highperformance)",
 "notes": "Corrects A2's gap note: HKU did build a complete catheter robot. dong2022_shape is CC BY 4.0 (printed on its p. 1), but its Fig. 6 slave photo carries labels and leader lines across the robot, so the picture is wang2018 Fig. 1. lee2018 counts 4 DoF (coarse + fine insertion), the 2018 IJCARS and 2022 papers 3.",
 "maintainers_own": false,
 "found_by": "gap"
}