 "schema_version": 1,
 "id": "hanglok_wang2024",
 "name": "Hanglok-Tech VIR",
 "full_name": "Hanglok-Tech vascular interventional robot (distributed architecture with DSA vision-servoing co-pilot)",
 "group": "Hanglok-Tech Co. Ltd., with Southeast University, South China University of Technology, Macau University of Science and Technology, Guangzhou First People's Hospital",
 "group_short": "Hanglok-Tech",
 "family": "academic",
 "category": "academic",
 "first_year": 2024,
 "latest_year": 2026,
 "versions": "wang2024_vision (CBS 2024, phantom in an interventional suite) -> zhu2026_bending (T-RO 2026, adds phantom paths and in vivo trials)",
 "target_anatomy": "head-and-neck / arch phantom (Elastrat); porcine vessels",
 "devices": [
  "guidewire",
  "microcatheter",
  "catheter",
  "sheath"
 ],
 "devices_driven": "0.014 in microwire, guide catheter (8 F sheath in phantom), microcatheter; contrast injection module",
 "n_dof": null,
 "n_dof_detail": "N.R.",
 "translation_mechanism": "other",
 "translation_detail": "other: microwire by an electric-gripper clamp on a lead-screw carriage with reverse drive (alternate clamp/release); microcatheter by dual friction wheels with multitooth Shore A70 silicone sleeve; guide catheter by advance-rotation module with Y-valve",
 "rotation_mechanism": "rotating-clamp",
 "rotation_detail": "Microwire twisted by the clamp ('guidewire twister', not friction wheels); guide catheter rotated through bevel gears and clutch",
 "device_range": "0.014 in microwire (nominal 0.36 mm); catheter sizes N.R.",
 "footprint_mm": "800x400x250 (jin2025_natural Sec. III-C)",
 "mass_kg": "N.R.",
 "force_sensing": "Pressure and torque sensors on USART ports of the embedded boards (zhu2026 Fig. 5); no force numbers reported in the sections read",
 "haptics_leader": "Hand-held controller with joystick and touch screen (no haptics described)",
 "imaging_compat": "Runs in a clinical angiography suite with DSA frames fed to the controller (X-ray)",
 "sterility": "Disinfected end effector; disposable contrast injector; further sterile-barrier detail N.R.",
 "highest_validation": "animal",
 "validation_detail": "Phantom: six vascular paths. Animal: nine in vivo vessel-navigation trials in pigs under IACUC approval of 20 Feb 2024 (zhu2026).",
 "key_numbers": "Vision co-pilot lowered the mean maximum bending-risk value 113.0 -> 49.0 in phantom (-56.6 %) and 239.8 -> 80.5 in animals (-66.4 %) (zhu2026 abstract)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "B",
  "kind": "photo",
  "source": "jin2025_natural Fig. 6(b) (prototype of the follower manipulator, MULLER, designed by Zhuhai Hanglok per Sec. III-C), T-HMS 2025, doi:10.1109/THMS.2025.3591542",
  "source_url": "https://doi.org/10.1109/THMS.2025.3591542",
  "licence": "publisher copyright",
  "credit": "Jin et al., IEEE Transactions on Human-Machine Systems (2025), Fig. 6(b). publisher copyright.",
  "edits": "native extract (2000x1931 figure); main cropped to the drive unit",
  "file": null
 },
 "pictures": [
  {
   "id": "hanglok_wang2024_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "MULLER follower on the angiography table beside the DSA device",
   "source_key": "jin2025_natural",
   "source": "Jin et al., IEEE Transactions on Human-Machine Systems (2025), Fig. 6(c), p. 5",
   "source_url": "https://doi.org/10.1109/thms.2025.3591542",
   "where": "Fig. 6(c), p. 5",
   "licence": "publisher copyright"
  },
  {
   "id": "hanglok_wang2024_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Robot in the intervention room during the Bama pig trial",
   "source_key": "zhu2026_bending",
   "source": "Zhu et al., IEEE Transactions on Robotics (2026), Fig. 9(b), p. 10",
   "source_url": "https://doi.org/10.1109/tro.2026.3712458",
   "where": "Fig. 9(b), p. 10",
   "licence": "publisher copyright"
  },
  {
   "id": "hanglok_wang2024_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Robot with the aortic phantom, control room and intervention room",
   "source_key": "zhu2026_bending",
   "source": "Zhu et al., IEEE Transactions on Robotics (2026), Fig. 9(a), p. 10",
   "source_url": "https://doi.org/10.1109/tro.2026.3712458",
   "where": "Fig. 9(a), p. 10",
   "licence": "publisher copyright"
  },
  {
   "id": "hanglok_wang2024_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Follower manipulator, haptic leader device and joystick on the bench",
   "source_key": "jin2025_natural",
   "source": "Jin et al., IEEE Transactions on Human-Machine Systems (2025), Fig. 12, p. 7",
   "source_url": "https://doi.org/10.1109/thms.2025.3591542",
   "where": "Fig. 12, p. 7",
   "licence": "publisher copyright"
  },
  {
   "id": "hanglok_wang2024_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Follower CAD with the disposable consumable lid open",
   "source_key": "jin2025_natural",
   "source": "Jin et al., IEEE Transactions on Human-Machine Systems (2025), Fig. 6(a), p. 5",
   "source_url": "https://doi.org/10.1109/thms.2025.3591542",
   "where": "Fig. 6(a), p. 5",
   "licence": "publisher copyright"
  },
  {
   "id": "hanglok_wang2024_06",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "System modules in the interventional operating room, robot on the table",
   "source_key": "wang2024_vision",
   "source": "Wang et al., 2024 IEEE International Conference on Cyborg and Bionic Systems (CBS) (2024), Fig. 5 (left), p. 5",
   "source_url": "https://doi.org/10.1109/cbs61689.2024.10860327",
   "where": "Fig. 5 (left), p. 5",
   "licence": "publisher copyright"
  },
  {
   "id": "hanglok_wang2024_07",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "CAD of the microwire advancement and guide-catheter modules",
   "source_key": "zhu2026_bending",
   "source": "Zhu et al., IEEE Transactions on Robotics (2026), Fig. 4(b)-(c), p. 5",
   "source_url": "https://doi.org/10.1109/tro.2026.3712458",
   "where": "Fig. 4(b)-(c), p. 5",
   "licence": "publisher copyright"
  },
  {
   "id": "hanglok_wang2024_08",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "Hand-held control unit with control stick, motor and circuit board",
   "source_key": "zhu2026_bending",
   "source": "Zhu et al., IEEE Transactions on Robotics (2026), Fig. 6, p. 6",
   "source_url": "https://doi.org/10.1109/tro.2026.3712458",
   "where": "Fig. 6, p. 6",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "wang2024_vision",
   "doi": "10.1109/cbs61689.2024.10860327",
   "url": null,
   "title": "Vision Servoing Endovascular Interventional Robot System: Design and Verification",
   "year": 2024,
   "venue": "2024 IEEE International Conference on Cyborg and Bionic Systems (CBS)"
  },
  {
   "key": "zhu2026_bending",
   "doi": "10.1109/tro.2026.3712458",
   "url": null,
   "title": "Bending-Aware Vision Co-Pilot for Intelligent Robotic Assistance in Endovascular Intervention",
   "year": 2026,
   "venue": "IEEE Transactions on Robotics"
  },
  {
   "key": "jin2025_natural",
   "doi": "10.1109/thms.2025.3591542",
   "url": null,
   "title": "Natural Human–Robot Interaction for Vascular Interventional Surgery: Design and Evaluation of a Leader Device With Haptic Feedback",
   "year": 2025,
   "venue": "IEEE Transactions on Human-Machine Systems"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (wang2024, zhu2026 text); abstract (jin2025)",
 "notes": "Hanglok is a company; commercial status was not established in the papers, so the row is filed as academic with a company-affiliated prototype. The perception module is a trained DSA segmentation network, but it was not trained in simulation. | 2026-10-13 coordinator: footprint from jin2025_natural Sec. III-C (MULLER follower; W4, P2)",
 "maintainers_own": false,
 "found_by": "academic"
}