 "schema_version": 1,
 "id": "dmci_peng2023",
 "name": "DMCI",
 "full_name": "DMCI: ergonomic dual-use mechanism (master interface or compact slave with unlimited travel) for cardiovascular intervention",
 "group": "Authors Peng, Wang, Xie, Gu (affiliation not retrieved; Lixu Gu is at Shanghai Jiao Tong University, not verified from the abstract)",
 "group_short": "Peng et al.",
 "family": "academic",
 "category": "academic",
 "first_year": 2022,
 "latest_year": 2023,
 "versions": "peng2023_design (IJCARS 2023)",
 "target_anatomy": "cardiovascular intervention (bench)",
 "devices": [
  "guidewire",
  "catheter"
 ],
 "devices_driven": "interventional tool (guidewire/catheter)",
 "n_dof": null,
 "n_dof_detail": "translation + rotation, motion decoupling and coupling kinematics",
 "translation_mechanism": "not-reported",
 "translation_detail": "N.R.",
 "rotation_mechanism": "not-reported",
 "rotation_detail": "N.R.",
 "device_range": "N.R.",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "N.R.",
 "haptics_leader": "DMCI doubles as the ergonomic leader interface; compared against a joystick",
 "imaging_compat": "N.R.",
 "sterility": "N.R.",
 "highest_validation": "phantom",
 "validation_detail": "Kinematic verification and master-slave teleoperation tests; operator study vs joystick (sizes N.R.)",
 "key_numbers": "Kinematic mean errors < 1 deg and < 1 mm; master-slave errors 0.821 +/- 0.753 deg and 0.608 +/- 0.512 mm",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "A",
  "kind": "photo",
  "source": "peng2023_design Fig. 2(a) (prototype adopting DMCI), IJCARS 2023, doi:10.1007/s11548-022-02755-4",
  "source_url": "https://doi.org/10.1007/s11548-022-02755-4",
  "licence": "publisher copyright",
  "credit": "Peng et al., International Journal of Computer Assisted Radiology and Surgery (2022), Fig. 2(a). publisher copyright.",
  "edits": "native extract (2032x1154 figure); cropped to panel (a); panel letter on the white background whitened; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "dmci_peng2023_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Teleoperation setup: camera, phantom, DMCI master, gamepad and slave",
   "source_key": "peng2023_design",
   "source": "Peng et al., International Journal of Computer Assisted Radiology and Surgery (2022), Fig. 7, p. 8",
   "source_url": "https://doi.org/10.1007/s11548-022-02755-4",
   "where": "Fig. 7, p. 8",
   "licence": "publisher copyright"
  },
  {
   "id": "dmci_peng2023_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Bench evaluation of the DMCI master and slave devices",
   "source_key": "peng2023_design",
   "source": "Peng et al., International Journal of Computer Assisted Radiology and Surgery (2022), Fig. 4, p. 6",
   "source_url": "https://doi.org/10.1007/s11548-022-02755-4",
   "where": "Fig. 4, p. 6",
   "licence": "publisher copyright"
  },
  {
   "id": "dmci_peng2023_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Internal motion of the DMCI: translation, rotation and both",
   "source_key": "peng2023_design",
   "source": "Peng et al., International Journal of Computer Assisted Radiology and Surgery (2022), Fig. 3, p. 4",
   "source_url": "https://doi.org/10.1007/s11548-022-02755-4",
   "where": "Fig. 3, p. 4",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "peng2023_design",
   "doi": "10.1007/s11548-022-02755-4",
   "url": null,
   "title": "Design, development and evaluation of an ergonomically designed dual-use mechanism for robot-assisted cardiovascular intervention",
   "year": 2022,
   "venue": "International Journal of Computer Assisted Radiology and Surgery"
  }
 ],
 "read_level": "abstract",
 "read_level_detail": "abstract (Europe PMC)",
 "notes": "Group must be identified from the PDF. | 2026-10-13 coordinator: highest_validation bench -> phantom from the full text (W4_notes.md)",
 "maintainers_own": false,
 "found_by": "academic"
}