 "schema_version": 1,
 "id": "cuhkmobileem_yang2021",
 "name": "Mobile coils + US",
 "full_name": "Three-electromagnet parallel mechanism with US probe for guidewire steering",
 "group": "Zhang, CUHK Mechanical and Automation Engineering",
 "group_short": "CUHK",
 "family": "magnetic",
 "category": "magnetic-navigation",
 "first_year": 2021,
 "latest_year": 2021,
 "versions": "single (RA-L 2021)",
 "target_anatomy": "abdominal/peripheral phantom (bifurcated 5 mm ID silicone channel)",
 "devices": [
  "guidewire"
 ],
 "devices_driven": "own replica-moulded guidewire: 50 um nitinol wire, PDMS cannula 1 mm OD, NdFeB magnet 1.2 mm OD x 2 mm at tip",
 "n_dof": null,
 "n_dof_detail": "parallel mechanism of three electromagnets and US probe; 1-axis motorised feeder (1-15 mm/s, devices 0.5-5 mm)",
 "translation_mechanism": "magnetic",
 "translation_detail": "magnetic (steering) + motorised feeder",
 "rotation_mechanism": "magnetic-field",
 "rotation_detail": "field direction",
 "device_range": "N.R.",
 "footprint_mm": "cylindrical workspace 500 mm diameter",
 "mass_kg": "N.R.",
 "force_sensing": "none",
 "haptics_leader": "no",
 "imaging_compat": "B-mode US at 22 Hz for tracking (radiation-free)",
 "sterility": "N.R.",
 "highest_validation": "phantom",
 "validation_detail": "three paths x three trials, tip reached target in all trials",
 "key_numbers": "30 mT per electromagnet at 100 mm, 12 mT omnidirectional at 100 mm; mean tip error 1.79 mm; speed 3.08 mm/s; kinematic modelling error 0.89 mm; field kept at 5 mT during run",
 "sim_model": "constant-curvature kinematic model with corrective angle (full)",
 "has_sim_model": true,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "B",
  "kind": "photo",
  "source": "yang2021_magnetic Fig. 8(b) (parallel mechanism with three electromagnets)",
  "source_url": "https://doi.org/10.1109/lra.2021.3057295",
  "licence": "publisher copyright (IEEE)",
  "credit": "Yang et al., IEEE Robotics and Automation Letters (2021), Fig. 8(b). publisher copyright (IEEE).",
  "edits": "native extract, crop",
  "file": null
 },
 "pictures": [
  {
   "id": "cuhkmobileem_yang2021_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Three electromagnets and ultrasound probe over the water-tank phantom with feeder",
   "source_key": "yang2021_magnetic",
   "source": "Yang et al., IEEE Robotics and Automation Letters (2021), Fig. 8(a), p. 7",
   "source_url": "https://doi.org/10.1109/lra.2021.3057295",
   "where": "Fig. 8(a), p. 7",
   "licence": "publisher copyright"
  },
  {
   "id": "cuhkmobileem_yang2021_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Concept rendering: parallel electromagnet mechanism with US probe above a patient",
   "source_key": "yang2021_magnetic",
   "source": "Yang et al., IEEE Robotics and Automation Letters (2021), Fig. 1, p. 2",
   "source_url": "https://doi.org/10.1109/lra.2021.3057295",
   "where": "Fig. 1, p. 2",
   "licence": "publisher copyright"
  },
  {
   "id": "cuhkmobileem_yang2021_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Guidewire tip deflection under rotating external fields, image series",
   "source_key": "yang2021_magnetic",
   "source": "Yang et al., IEEE Robotics and Automation Letters (2021), Fig. 7, p. 6",
   "source_url": "https://doi.org/10.1109/lra.2021.3057295",
   "where": "Fig. 7, p. 6",
   "licence": "publisher copyright"
  },
  {
   "id": "cuhkmobileem_yang2021_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Structure of the magnetic guidewire: magnet segment, spring and core",
   "source_key": "yang2021_magnetic",
   "source": "Yang et al., IEEE Robotics and Automation Letters (2021), Fig. 2(a), p. 2",
   "source_url": "https://doi.org/10.1109/lra.2021.3057295",
   "where": "Fig. 2(a), p. 2",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "yang2021_magnetic",
   "doi": "10.1109/lra.2021.3057295",
   "url": null,
   "title": "Magnetic Control of a Steerable Guidewire Under Ultrasound Guidance Using Mobile Electromagnets",
   "year": 2021,
   "venue": "IEEE Robotics and Automation Letters"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full",
 "notes": "Crop clips panel (b) at its right edge.",
 "maintainers_own": false,
 "found_by": "actuated-tip"
}