 "schema_version": 1,
 "id": "coast_jeong2020",
 "name": "COAST",
 "full_name": "COAST guidewire robot (coaxially aligned steerable)",
 "group": "Desai, Georgia Tech Robomed Lab",
 "group_short": "Georgia Tech",
 "family": "tip",
 "category": "actuated-device",
 "first_year": 2020,
 "latest_year": 2025,
 "versions": "2020 3-tube design 0.40 mm OD (jeong2020_design); 2022 same family run under fluoroscopic path planning, ex vivo porcine aorta (ravigopal2022_fluoroscopic); 2023 larger 0.89 mm OD with 3-fibre FBG shape+force sensing (deaton2023_simultaneous); 2025 two-tube 'modified COAST', 0.35 mm outer tube, 0.66 mm Pebax sheath, joystick teleoperation (brumfiel2025_modified)",
 "target_anatomy": "cardiovascular / peripheral (aortic bifurcation phantom, paediatric carotid and arch phantoms)",
 "devices": [
  "guidewire"
 ],
 "devices_driven": "the COAST wire itself (own nitinol-tube wire, not a commercial guidewire)",
 "n_dof": 4,
 "n_dof_detail": "4 actuated axes in 2020 prototype (tendon stroke X1, inner-tube position X2, outer-tube advance X3, whole-module advance X4); roll added by spur gear in 2025 version",
 "translation_mechanism": "tendon/tip-actuation",
 "translation_detail": "tendon/tip-actuation",
 "rotation_mechanism": "rotating-drive-unit",
 "rotation_detail": "none in 2020 prototype; spur-gear roll of actuation module (2025)",
 "device_range": "2020 prototype 188-257 mm tubes, 'shorter than a clinical wire'; handheld spooling variant takes up to 1.5 m (chavenet2026_handheld)",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "2020: miniature force sensor on tendon tension; 2023: FBG triplet tip force, RMSE 0.005-0.027 N (bench)",
 "haptics_leader": "no (Xbox controller teleoperation 2025)",
 "imaging_compat": "fluoroscopy used 2022, 2025; MR compatibility N.R.",
 "sterility": "N.R.",
 "highest_validation": "ex-vivo",
 "validation_detail": "ex vivo porcine aorta, navigation with COAST under fluoroscopic reconstruction (ravigopal2022 abstract); 2025: one qualitative teleoperated traversal of an aortic-bifurcation phantom with pulsatile flow, no success counts; 2020: phantom, no success count",
 "key_numbers": "2020: tendon-stroke model RMSE 0.032 mm (27-37 mm bending lengths); follow-the-leader tip-to-path mean 4.53 mm (low curvature) and 14.66 mm (high curvature) in free space; 2023 shape RMSE 0.62-1.94 mm by motion type; 2025 wire buckled between introducer and actuator so insertion was hand-assisted",
 "sim_model": "analytical tendon-stroke / static model with experimental fit (jeong2020 Sec. III, full); shape-from-FBG reconstruction (deaton2023, full); no physics simulator found",
 "has_sim_model": true,
 "sim2real": "none-found",
 "sim2real_detail": "none found (all control open-loop or teleoperated; no learned policy)",
 "picture": {
  "status": "not-cleared",
  "grade": "A",
  "kind": "photo",
  "source": "deaton2023_simultaneous Fig. 1 (lower photo: COAST actuation stage with guidewire)",
  "source_url": "https://doi.org/10.1109/lra.2023.3267008",
  "licence": "publisher copyright (IEEE)",
  "credit": "Deaton et al., IEEE Robotics and Automation Letters (2023), Fig. 1. publisher copyright (IEEE).",
  "edits": "native extract (2014 px); cropped; 2 black arrowheads on the white background erased (LaMa, 3 passes, looked at 4x zoom)",
  "file": null
 },
 "pictures": [
  {
   "id": "coast_jeong2020_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "photo",
   "caption": "Side view of the four-carriage linear actuation stage on acrylic base",
   "source_key": "jeong2020_design",
   "source": "Jeong et al., IEEE Robotics and Automation Letters (2020), Fig. 3(c), p. 3",
   "source_url": "https://doi.org/10.1109/lra.2020.3004782",
   "where": "Fig. 3(c), p. 3",
   "licence": "publisher copyright"
  },
  {
   "id": "coast_jeong2020_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Modified COAST actuation unit on a robot arm beside the C-arm",
   "source_key": "brumfiel2025_modified",
   "source": "Brumfiel et al. (2025), Fig. 2(a-1), p. 2",
   "source_url": "https://arxiv.org/abs/2512.13477",
   "where": "Fig. 2(a-1), p. 2",
   "licence": "arXiv non-exclusive distribution licence 1.0"
  },
  {
   "id": "coast_jeong2020_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "COAST guidewire actuation system with ex vivo porcine aorta under C-arm",
   "source_key": "ravigopal2022_fluoroscopic",
   "source": "Ravigopal et al., IEEE Robotics and Automation Letters (2022), Fig. 9(a), p. 7",
   "source_url": "https://doi.org/10.1109/lra.2022.3207568",
   "where": "Fig. 9(a), p. 7",
   "licence": "publisher copyright"
  },
  {
   "id": "coast_jeong2020_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Rendering of the enclosed two-tube modified COAST actuation unit",
   "source_key": "brumfiel2025_modified",
   "source": "Brumfiel et al. (2025), Fig. 1, p. 1",
   "source_url": "https://arxiv.org/abs/2512.13477",
   "where": "Fig. 1, p. 1",
   "licence": "arXiv non-exclusive distribution licence 1.0"
  },
  {
   "id": "coast_jeong2020_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Rendering of the modified actuation unit with its cover removed",
   "source_key": "brumfiel2025_modified",
   "source": "Brumfiel et al. (2025), Fig. 1, p. 1",
   "source_url": "https://arxiv.org/abs/2512.13477",
   "where": "Fig. 1, p. 1",
   "licence": "arXiv non-exclusive distribution licence 1.0"
  },
  {
   "id": "coast_jeong2020_06",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "COAST guidewire choosing branches in a bifurcation phantom, image sequence",
   "source_key": "jeong2020_design",
   "source": "Jeong et al., IEEE Robotics and Automation Letters (2020), Fig. 9, p. 7",
   "source_url": "https://doi.org/10.1109/lra.2020.3004782",
   "where": "Fig. 9, p. 7",
   "licence": "publisher copyright"
  },
  {
   "id": "coast_jeong2020_07",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Notched nitinol tubes of the COAST guidewire under magnification",
   "source_key": "jeong2020_design",
   "source": "Jeong et al., IEEE Robotics and Automation Letters (2020), Fig. 3(a), p. 3",
   "source_url": "https://doi.org/10.1109/lra.2020.3004782",
   "where": "Fig. 3(a), p. 3",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "jeong2020_design",
   "doi": "10.1109/lra.2020.3004782",
   "url": null,
   "title": "Design, Modeling, and Control of a Coaxially Aligned Steerable (COAST) Guidewire Robot",
   "year": 2020,
   "venue": "IEEE Robotics and Automation Letters"
  },
  {
   "key": "deaton2023_simultaneous",
   "doi": "10.1109/lra.2023.3267008",
   "url": null,
   "title": "Simultaneous Shape and Tip Force Sensing for the COAST Guidewire Robot",
   "year": 2023,
   "venue": "IEEE Robotics and Automation Letters"
  },
  {
   "key": "ravigopal2022_fluoroscopic",
   "doi": "10.1109/lra.2022.3207568",
   "url": null,
   "title": "Fluoroscopic Image-Based 3-D Environment Reconstruction and Automated Path Planning for a Robotically Steerable Guidewire",
   "year": 2022,
   "venue": "IEEE Robotics and Automation Letters"
  },
  {
   "key": "brumfiel2025_modified",
   "doi": null,
   "url": "https://arxiv.org/abs/2512.13477",
   "title": "Evaluating the Navigation Capabilities of a Modified COAST Guidewire Robot in an Anatomical Phantom Model",
   "year": 2025,
   "venue": null
  },
  {
   "key": "chavenet2026_handheld",
   "doi": null,
   "url": "https://arxiv.org/abs/2607.24629",
   "title": "Development of a Handheld Actuation Mechanism for a Tendon-driven Robotically Steered Guidewire",
   "year": 2026,
   "venue": null
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (jeong2020, deaton2023, brumfiel2025 arXiv, chavenet2026 arXiv); abstract (ravigopal2022)",
 "notes": "Family has the most hardware-complete published tendon guidewire in my scope; the 2023 paper and the 2025 modified version change OD, so quote the version with every size.",
 "maintainers_own": false,
 "found_by": "actuated-tip"
}