 "schema_version": 1,
 "id": "chonnam_jeong2017",
 "name": "PCI magnetic wire",
 "full_name": "Magnetically steerable PCI guidewire with a Helmholtz-coil EMA and a motorized feeding stage",
 "group": "J.-O. Park, S. Park, C.-S. Kim, Chonnam National University and Korea Medical Microrobot Center",
 "group_short": "Chonnam NU",
 "family": "magnetic",
 "category": "magnetic-navigation",
 "first_year": 2017,
 "latest_year": 2019,
 "versions": "2017: two pairs of Helmholtz coils + motorized feeding stage, feed-forward + PID steering (jeong2017_feasibility); 2019: EMA of two Helmholtz pairs and one Maxwell pair (one x-axis pair used) + linear-stage feeder, path regeneration, feeding control and vision tracking (kim2019_novel); 2018-21: same group's guide-wired helical microrobot (GHM) for mechanical atherectomy: 3D-printed corkscrew head (1.6 x 9.5 mm, NdFeB magnet magnetised across the axis) on a spherical joint at the tip of a 0.152 mm nitinol guidewire, steered by a uniform field and spun for drilling by a rotating field (18-370 Hz, resonant circuits) of three Helmholtz-coil pairs (30 mT), advanced and retracted by a motorised linear-stage feeder in an automatic tracing mode; in vitro feasibility (Nguyen 2018 EMBC); phantom path tracking, gelatin-thrombus drilling and 5 rats in vivo (Nguyen 2021 IEEE TBME 68:2490-2498, Secs. II-IV, Figs. 2, 4) (Konda 2025 Table 2 row 8) (nguyen2021_guidewired; konda2025_robotically)",
 "target_anatomy": "coronary (PCI), tested in planar vessel phantoms; 2021: chronic total occlusion; rat abdominal aorta used as a coronary-size model (nguyen2021_guidewired Sec. IV-G)",
 "devices": [
  "guidewire",
  "catheter",
  "sheath"
 ],
 "devices_driven": "2017: Asahi Conquest Pro 12 guidewire (tip OD 0.23 mm) with a bullet-shaped NdFeB magnet 1 mm x 7 mm (jeong2017_feasibility Sec. 2.1); 2019: 0.2 mm nitinol wire with a magnet tip (kim2019_novel Sec. 4.1); 2021: 0.152 mm super-elastic nitinol guidewire (Confluent Medical) carrying the 3D-printed helical head; 130 mm of wire through a Tygon sheath (0.79 mm ID) as guiding catheter (nguyen2021_guidewired Secs. II-A, IV-A)",
 "n_dof": 2,
 "n_dof_detail": "2: insertion (feeder) + planar tip steering (field); 2021: insertion (feeder) + 3-D steering and spin of the head (field)",
 "translation_mechanism": "linear-stage+clamp",
 "translation_detail": "linear-stage+clamp (motorized feeding stage with a guidewire holder, jeong2017_feasibility abstract; linear stage, kim2019_novel Fig. 1); tip steered magnetically",
 "rotation_mechanism": "magnetic-field",
 "rotation_detail": "field direction of the Helmholtz coils (no shaft rotation); 2021: a rotating field spins the helical head on its spherical joint; the guidewire does not turn (nguyen2021_guidewired Secs. II-A, III)",
 "device_range": "N.R.",
 "footprint_mm": "Helmholtz coil radius 162 mm, region of interest 100 x 100 (kim2019_novel Sec. 2.1)",
 "mass_kg": "N.R.",
 "force_sensing": "none",
 "haptics_leader": "no",
 "imaging_compat": "overhead CCD camera only (jeong2017_feasibility Sec. 3.1; kim2019_novel); 2021: optical camera on the phantom, C-arm X-ray (Ziehm Vision RFD) in the rat study (nguyen2021_guidewired Sec. IV-A)",
 "sterility": "N.R.",
 "highest_validation": "animal",
 "validation_detail": "2017: five-branch phantom (about 4 mm branches), insertion success 100% with feed-forward + PID vs about 75% with feed-forward only (jeong2017_feasibility Sec. 3.3); 2019: smooth and sharply curved vessel phantoms, 3 paths x 3 replicates (kim2019_novel Sec. 4, Table 1); 2021: 5 rats in vivo (abdominal aorta, injected 9 % gelatin thrombus, contrast angiography): thrombus drilled without perforation in 3, blood flow restored in 2, rats 1 and 2 died during thrombus formation and rat 3 of bleeding after drilling; vessel phantom path tracking RMSE about 0.2 mm, maximum below 0.7 mm at 0.3 mm/s; gelatin thrombi in tubes drilled through, 54 mm (9 %) in about 48 s and 40 mm (11 %) in 43 s (nguyen2021_guidewired Secs. IV-D to IV-G, Table II, Figs. 6-8)",
 "key_numbers": "steering-angle error below 0.5 deg with PID vs about 7 deg feed-forward only (jeong2017_feasibility Sec. 3.2); tip-position RMSE 0.44 +/- 0.19 mm on path I, below 0.7 +/- 0.4 mm on all paths; feeding speed 1 mm/s; camera 0.1 mm per pixel (kim2019_novel Sec. 3-4); 2021: three Helmholtz pairs on three 15 kW supplies, 30 mT, rotating field up to 370 Hz; drilling force up to 0.8 N and torque 0.1125 mN m at 100 Hz; feedback steering error below 0.5 deg (maximum 0.52 deg) against about 3 deg feed-forward (nguyen2021_guidewired Secs. III, IV-A, IV-D)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "C",
  "kind": "photo",
  "source": "kim2019_novel Fig. 1",
  "source_url": "https://doi.org/10.1007/s12555-018-0116-8",
  "licence": "publisher copyright",
  "credit": "Kim et al., International Journal of Control, Automation and Systems (2019), Fig. 1. publisher copyright.",
  "edits": "native extract from kim2019_tip-positioning_IJCAS.pdf p. 3; no crop; none upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "chonnam_jeong2017_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Helmholtz coils, camera, guidewire device and motorized feeding stage",
   "source_key": "jeong2017_feasibility",
   "source": "Jeong et al., International Journal of Control, Automation and Systems (2016), Fig. 2(b), p. 3",
   "source_url": "https://doi.org/10.1007/s12555-015-0269-7",
   "where": "Fig. 2(b), p. 3",
   "licence": "publisher copyright"
  },
  {
   "id": "chonnam_jeong2017_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Helmholtz-coil EMA, feeding device, resonant circuit and C-arm in the shielded room",
   "source_key": "nguyen2021_guidewired",
   "source": "Nguyen et al., IEEE Transactions on Biomedical Engineering (2021), Fig. 4(a), p. 2493",
   "source_url": "https://doi.org/10.1109/tbme.2020.3046513",
   "where": "Fig. 4(a), p. 2493",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "chonnam_jeong2017_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "3D-printed helical microrobot head on the end of the guidewire",
   "source_key": "nguyen2021_guidewired",
   "source": "Nguyen et al., IEEE Transactions on Biomedical Engineering (2021), Fig. 2(b), p. 2491",
   "source_url": "https://doi.org/10.1109/tbme.2020.3046513",
   "where": "Fig. 2(b), p. 2491",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "chonnam_jeong2017_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Helical head drilling through a 54 mm gelatin thrombus in a tube over 48 s",
   "source_key": "nguyen2021_guidewired",
   "source": "Nguyen et al., IEEE Transactions on Biomedical Engineering (2021), Fig. 6(c), p. 2494",
   "source_url": "https://doi.org/10.1109/tbme.2020.3046513",
   "where": "Fig. 6(c), p. 2494",
   "licence": "publisher copyright (IEEE)"
  }
 ],
 "sources": [
  {
   "key": "jeong2017_feasibility",
   "doi": "10.1007/s12555-015-0269-7",
   "url": null,
   "title": "Feasibility study on magnetically steerable guidewire device for percutaneous coronary intervention",
   "year": 2016,
   "venue": "International Journal of Control, Automation and Systems"
  },
  {
   "key": "kim2019_novel",
   "doi": "10.1007/s12555-018-0116-8",
   "url": null,
   "title": "A Novel Tip-positioning Control of a Magnetically Steerable Guidewire in Sharply Curved Blood Vessel for Percutaneous Coronary Intervention",
   "year": 2019,
   "venue": "International Journal of Control, Automation and Systems"
  },
  {
   "key": "konda2025_robotically",
   "doi": "10.1126/scirobotics.adt7461",
   "url": null,
   "title": "Robotically steerable guidewires—Current trends and future directions",
   "year": 2025,
   "venue": "Science Robotics"
  },
  {
   "key": "nguyen2021_guidewired",
   "doi": "10.1109/tbme.2020.3046513",
   "url": null,
   "title": "Guide-Wired Helical Microrobot for Percutaneous Revascularization in Chronic Total Occlusion <i>in-Vivo</i> Validation",
   "year": 2021,
   "venue": "IEEE Transactions on Biomedical Engineering"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (jeong2017_feasibility, kim2019_novel, nguyen2021_guidewired)",
 "notes": "Family: magnetic. konda2025 Table 2 rows 5 and 7. The same group's guide-wired helical microrobot (Nguyen 2021 TBME, konda2025 row 8) was judged 'unclear' in review_coverage and is not added. | N4 2026-11-09: nguyen2021_guidewired read in full (the 'not added' sentence above predates N3, which filed the helical microrobot as a version line): same group's three-pair Helmholtz EMA with a motorised linear-stage feeder; 5 rats in vivo, so highest_validation phantom -> animal; C-arm X-ray added to imaging; first_year and latest_year kept at 2017 and 2019, as for every round-8 version line",
 "maintainers_own": false,
 "found_by": "actuated-tip"
}