 "schema_version": 1,
 "id": "cathrob_cercenelli2017",
 "name": "CathROB",
 "full_name": "CathROB remote catheter navigation system (with its 2007-2008 telerobotic predecessor)",
 "group": "University of Bologna, Laboratory of Bioengineering (L. Cercenelli, E. Marcelli, G. Plicchi, B. Bortolani), developed with Tre Esse Progettazione Biomedica",
 "group_short": "U. Bologna",
 "family": "academic",
 "category": "academic",
 "first_year": 2007,
 "latest_year": 2017,
 "versions": "2007 ASAIO J telerobotic system (cercenelli2007_initial, full): driving unit on an articulation arm giving advance/withdrawal, rotation and tip steering to a standard EP catheter, PIC16F876 control unit at the bedside, remote push-button box with display (wired or wireless), automatic repositioning to stored targets (Methods, Figs. 1-2); 2008 Computers in Cardiology telerobotic system with force sensing (title); 2017 CathROB: mock handle plus push-button box, compact arm-mounted Motion Unit (full)",
 "target_anatomy": "cardiac electrophysiology mapping and RF ablation (right atrium in sheep; silicone left-atrium model)",
 "devices": [
  "catheter"
 ],
 "devices_driven": "one standard tip-steerable EP catheter; adapters for Navistar Thermocool and EZ-steer bidirectional (Biosense Webster), Sprinklr XL 7 Fr (Medtronic), Blazer II XP (Boston Scientific); tested with Navistar Thermocool 4 mm 7.5 Fr (Sec. 2.1); 2007: RF Enhancr steerable EP catheter (Medtronic) (cercenelli2007 Methods)",
 "n_dof": 3,
 "n_dof_detail": "3 (advance/withdraw, rotation, tip bend/straighten)",
 "translation_mechanism": "linear-stage+clamp",
 "translation_detail": "linear-stage+clamp (Motion Unit moves the handle held by a spring-preloaded jaw block with silicone-coated rollers; Delrin telescopic sheath to the introducer)",
 "rotation_mechanism": "handle-rotation",
 "rotation_detail": "DC micromotor rotates the gripped handle; steering element turned through catheter-specific adapters",
 "device_range": "four commercial EP catheter models via adapters",
 "footprint_mm": "Motion Unit 53 x 8 x 11 cm (Fig. 4 labels; text prints 53 x 8 x 110 cm) retracted, 73 x 8 x 11 cm extended (Sec. 2.1)",
 "mass_kg": "2.3 (Motion Unit); supporting arm 7.5 (Sec. 2.1, Fig. 4)",
 "force_sensing": "Honeywell FS-series piezoresistive sensor in the push path; alarm at 60% and stop at 80% of a perforation force of 128 +- 12 gF measured on excised ovine atrium (Sec. 2.1)",
 "haptics_leader": "mock catheter handle plus push-button box; vibration motor and acoustic alarm when thresholds are exceeded; up to four stored targets for automatic repositioning",
 "imaging_compat": "standard fluoroscopy and CARTO electroanatomic mapping; operator outside the radiation field",
 "sterility": "disposable or sterilisable adapters; Motion Unit and arm draped in sterile polyethylene in vivo",
 "highest_validation": "animal",
 "validation_detail": "in vitro silicone heart and vascular mock-up; in vivo 4 sheep (49 +- 4 kg) with right-atrial CARTO mapping and RF ablation at RA isthmus and posterior wall (Sec. 2.2, 3); 2007 predecessor: 3 sheep (58 +- 3 kg), navigation and repositioning to 12 predefined endocardial targets under fluoroscopy, manual vs robotic (cercenelli2007 Methods, Results)",
 "key_numbers": "automatic repositioning error 1.5 +- 0.6 mm (left targets) and 1.7 +- 0.5 mm (right; abstract 1.7 +- 0.4); map-to-CT deviation 1.37 +- 0.98 mm; in vivo set-up 5 +- 3 min, maps 63 +- 8 points in 36 +- 7 min with 15 +- 3 min fluoroscopy; speeds 13-18 mm/s, 30-45 deg/s, 2-4 mm/s; no perforation or rhythm change (Sec. 2.1, 3); 2007: navigation 13.0 +- 5.6 s robotic vs 16.1 +- 6.4 s manual, repositioning 9.2 +- 2.5 vs 17.8 +- 7.1 s (both p < 0.001); bench repositioning accuracy 0.2 +- 0.1 mm, 1.4 +- 0.8 mm, 2.1 +- 0.9 mm (longitudinal, rotational, steering) (cercenelli2007 Abstract, Methods)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "cleared",
  "grade": "B",
  "kind": "photo",
  "source": "cercenelli2017_cathrob Fig. 2(b) top (Motion Unit on articulated arm), native embedded image on PDF p. 3 from the PMC open-access PDF",
  "source_url": "https://doi.org/10.1155/2017/2712453",
  "licence": "CC BY 4.0",
  "credit": "Cercenelli et al., Applied Bionics and Biomechanics (2017), Fig. 2(b). CC BY 4.0.",
  "edits": "native extract (1586 x 1320); cropped to Motion Unit and upper arm; no erasing (labels were vector overlays outside the bitmap); not upscaled (upscale disabled)",
  "file": "pictures/cathrob_cercenelli2017.jpg"
 },
 "pictures": [
  {
   "id": "cathrob_cercenelli2017_01",
   "status": "cleared",
   "file": "pictures/extra/cathrob_cercenelli2017_01.jpg",
   "thumb": "pictures/thumbs/cathrob_cercenelli2017_01.jpg",
   "kind": "in-use",
   "caption": "Draped Motion Unit on its arm prepared for the in-vivo animal study",
   "source_key": "cercenelli2017_cathrob",
   "source": "Cercenelli et al., Applied Bionics and Biomechanics (2017), Fig. 9, p. 9",
   "source_url": "https://doi.org/10.1155/2017/2712453",
   "where": "Fig. 9, p. 9",
   "licence": "CC BY 4.0 (Hindawi; PDF states 'Creative Commons Attribution License')"
  },
  {
   "id": "cathrob_cercenelli2017_02",
   "status": "cleared",
   "file": "pictures/extra/cathrob_cercenelli2017_02.jpg",
   "thumb": "pictures/thumbs/cathrob_cercenelli2017_02.jpg",
   "kind": "in-use",
   "caption": "In-vitro set-up: arm-mounted Motion Unit feeding a catheter into a vascular model",
   "source_key": "cercenelli2017_cathrob",
   "source": "Cercenelli et al., Applied Bionics and Biomechanics (2017), Fig. 8(a), p. 8",
   "source_url": "https://doi.org/10.1155/2017/2712453",
   "where": "Fig. 8(a), p. 8",
   "licence": "CC BY 4.0 (Hindawi; PDF states 'Creative Commons Attribution License')"
  },
  {
   "id": "cathrob_cercenelli2017_03",
   "status": "cleared",
   "file": "pictures/extra/cathrob_cercenelli2017_03.jpg",
   "thumb": "pictures/thumbs/cathrob_cercenelli2017_03.jpg",
   "kind": "setup",
   "caption": "Whole system: arm-mounted Motion Unit, GUI monitor, control box and push-button box",
   "source_key": "cercenelli2017_cathrob",
   "source": "Cercenelli et al., Applied Bionics and Biomechanics (2017), Fig. 2(b), p. 3",
   "source_url": "https://doi.org/10.1155/2017/2712453",
   "where": "Fig. 2(b), p. 3",
   "licence": "CC BY 4.0 (Hindawi; PDF states 'Creative Commons Attribution License')"
  },
  {
   "id": "cathrob_cercenelli2017_04",
   "status": "cleared",
   "file": "pictures/extra/cathrob_cercenelli2017_04.jpg",
   "thumb": "pictures/thumbs/cathrob_cercenelli2017_04.jpg",
   "kind": "close-up",
   "caption": "Motion Unit holding an EP catheter handle, with clamp insets",
   "source_key": "cercenelli2017_cathrob",
   "source": "Cercenelli et al., Applied Bionics and Biomechanics (2017), Fig. 3, p. 4",
   "source_url": "https://doi.org/10.1155/2017/2712453",
   "where": "Fig. 3, p. 4",
   "licence": "CC BY 4.0 (Hindawi; PDF states 'Creative Commons Attribution License')"
  },
  {
   "id": "cathrob_cercenelli2017_05",
   "status": "cleared",
   "file": "pictures/extra/cathrob_cercenelli2017_05.jpg",
   "thumb": "pictures/thumbs/cathrob_cercenelli2017_05.jpg",
   "kind": "photo",
   "caption": "Disassembled Motion Unit, supporting arm, anchoring plate and cable",
   "source_key": "cercenelli2017_cathrob",
   "source": "Cercenelli et al., Applied Bionics and Biomechanics (2017), Fig. 4, p. 4",
   "source_url": "https://doi.org/10.1155/2017/2712453",
   "where": "Fig. 4, p. 4",
   "licence": "CC BY 4.0 (Hindawi; PDF states 'Creative Commons Attribution License')"
  },
  {
   "id": "cathrob_cercenelli2017_06",
   "status": "cleared",
   "file": "pictures/extra/cathrob_cercenelli2017_06.jpg",
   "thumb": "pictures/thumbs/cathrob_cercenelli2017_06.jpg",
   "kind": "close-up",
   "caption": "Handle adapters fitting four commercial tip-steerable EP catheters",
   "source_key": "cercenelli2017_cathrob",
   "source": "Cercenelli et al., Applied Bionics and Biomechanics (2017), Fig. 5, p. 5",
   "source_url": "https://doi.org/10.1155/2017/2712453",
   "where": "Fig. 5, p. 5",
   "licence": "CC BY 4.0 (Hindawi; PDF states 'Creative Commons Attribution License')"
  },
  {
   "id": "cathrob_cercenelli2017_07",
   "status": "cleared",
   "file": "pictures/extra/cathrob_cercenelli2017_07.jpg",
   "thumb": "pictures/thumbs/cathrob_cercenelli2017_07.jpg",
   "kind": "leader",
   "caption": "Command interface: mock catheter handle and push-button box",
   "source_key": "cercenelli2017_cathrob",
   "source": "Cercenelli et al., Applied Bionics and Biomechanics (2017), Fig. 6, p. 6",
   "source_url": "https://doi.org/10.1155/2017/2712453",
   "where": "Fig. 6, p. 6",
   "licence": "CC BY 4.0 (Hindawi; PDF states 'Creative Commons Attribution License')"
  },
  {
   "id": "cathrob_cercenelli2017_08",
   "status": "cleared",
   "file": "pictures/extra/cathrob_cercenelli2017_08.jpg",
   "thumb": "pictures/thumbs/cathrob_cercenelli2017_08.jpg",
   "kind": "render",
   "caption": "3D rendering of the system attached to a cath-lab table by an articulated arm",
   "source_key": "cercenelli2017_cathrob",
   "source": "Cercenelli et al., Applied Bionics and Biomechanics (2017), Fig. 1, p. 2",
   "source_url": "https://doi.org/10.1155/2017/2712453",
   "where": "Fig. 1, p. 2",
   "licence": "CC BY 4.0 (Hindawi; PDF states 'Creative Commons Attribution License')"
  }
 ],
 "sources": [
  {
   "key": "cercenelli2017_cathrob",
   "doi": "10.1155/2017/2712453",
   "url": null,
   "title": "CathROB: A Highly Compact and Versatile Remote Catheter Navigation System",
   "year": 2017,
   "venue": "Applied Bionics and Biomechanics"
  },
  {
   "key": "cercenelli2007_initial",
   "doi": "10.1097/mat.0b013e3181484cd0",
   "url": null,
   "title": "Initial Experience With a Telerobotic System to Remotely Navigate and Automatically Reposition Standard Steerable EP Catheters",
   "year": 2007,
   "venue": "ASAIO Journal"
  },
  {
   "key": "marcelli2008_novel",
   "doi": "10.1109/cic.2008.4748996",
   "url": null,
   "title": "A novel telerobotic system to remotely navigate standard electrophysiology catheters",
   "year": 2008,
   "venue": "2008 Computers in Cardiology"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (CathROB 2017; 2007 ASAIO, ); title (2008)",
 "notes": "Venue is Applied Bionics and Biomechanics (Hindawi, CC BY 4.0). Only CathROB is listed by li2024_design as an academic reference system next to commercial ones. | 2026-11-08 X3: 2007 paper read in full. Its two photographs (Fig. 2 driving unit on the articulation arm; Fig. 3 in vivo sheep set-up with the operator at the push-button box, 975 px greyscale) are not added because the gallery already holds 8 pictures; they are candidates if the coordinator swaps one",
 "maintainers_own": false,