 "schema_version": 1,
 "id": "cathpilot_zhou2022",
 "name": "CathPilot",
 "full_name": "CathPilot (expandable cable-driven tip platform steered through a 3-D cam handle)",
 "group": "Toronto Metropolitan University (M. A. Tavallaei) with Sunnybrook Research Institute (G. A. Wright), University of Toronto (A. Dueck), Magellan Biomedical, University of Waterloo",
 "group_short": "Toronto Metropolitan",
 "family": "tip",
 "category": "actuated-device",
 "first_year": 2022,
 "latest_year": 2023,
 "versions": "2022 TMECH design and bench verification; 2023 TBME performance validation, ex-vivo CTO crossing and in-vivo porcine safety",
 "target_anatomy": "peripheral arteries, chronic total occlusion (CTO) crossing",
 "devices": [
  "guidewire",
  "catheter",
  "sheath"
 ],
 "devices_driven": "CathPilot catheter positions a standard guidewire tip; frame deliverable through a 12 Fr sheath (OD 5.3 mm), expanding to 20 mm (TMECH 2022 Sec. III)",
 "n_dof": 2,
 "n_dof_detail": "2 (planar positioning of the guidewire end-effector within the expanded frame)",
 "translation_mechanism": "tendon/tip-actuation",
 "translation_detail": "tendon/tip-actuation (four PTFE cables, 0.012 in, from handle to an end-effector suspended in an expandable frame)",
 "rotation_mechanism": "not-reported",
 "rotation_detail": "N.R. (planar 2-DOF)",
 "device_range": "standard guidewire through the end-effector; several frame opening diameters tested",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "none (passive)",
 "haptics_leader": "handle with a 2-DOF 3-D cam surface sets all four cable lengths: fully manual mechanical teleoperation, no motors (TMECH 2022 Sec. III)",
 "imaging_compat": "X-ray fluoroscopy (N.R. further)",
 "sterility": "N.R.",
 "highest_validation": "animal",
 "validation_detail": "TBME 2023: tortuous vessel phantom (target reach vs non-steerable and steerable catheters), ex-vivo human CTO samples (fresh and fixed), in-vivo porcine aorta in one male pig of 56.6 kg",
 "key_numbers": "target-reach success 31% non-steerable, 69% steerable, 100% CathPilot; CTO crossing 83% fresh and 100% fixed; up to four times higher force delivery and pushability (TBME 2023 abstract); submillimetre positioning (TMECH 2022 abstract)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "B",
  "kind": "render",
  "source": "zhou2022_cathpilot_TMECH Fig. 1 (CAD of manipulator unit, multilumen catheter, user-input handle)",
  "source_url": "https://doi.org/10.1109/tmech.2022.3188955",
  "licence": "CC BY-NC-ND 4.0 (article licence line); zhou2023_cathpilot_TBME is CC BY 4.0",
  "credit": "Zhou et al., IEEE/ASME Transactions on Mechatronics (2022), Fig. 1. CC BY-NC-ND 4.0 (article licence line); zhou2023_cathpilot_TBME is CC BY 4.0.",
  "edits": "native extract (2000x595); text labels (OCR) and 3 blue arrows on white background erased (LaMa); faint grey residue in the erased label areas set to white; top 95 px cropped; 40 px white padding (render background); not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "cathpilot_zhou2022_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "photo",
   "caption": "Handle prototype with thumbstick and tensioner lock on a 12 Fr sheath",
   "source_key": "zhou2022_cathpilot",
   "source": "Zhou et al., IEEE/ASME Transactions on Mechatronics (2022), Fig. 7, p. 6",
   "source_url": "https://doi.org/10.1109/tmech.2022.3188955",
   "where": "Fig. 7, p. 6",
   "licence": "publisher copyright"
  },
  {
   "id": "cathpilot_zhou2022_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "CathPilot with silicone tube and lesion fixture in the catheterization lab",
   "source_key": "zhou2023_cathpilot",
   "source": "Zhou et al., IEEE Transactions on Biomedical Engineering (2023), Fig. 2, p. 3",
   "source_url": "https://doi.org/10.1109/tbme.2023.3276199",
   "where": "Fig. 2, p. 3",
   "licence": "publisher copyright"
  },
  {
   "id": "cathpilot_zhou2022_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Expandable cable-driven frame emerging from the delivery sheath, four stages",
   "source_key": "zhou2022_cathpilot",
   "source": "Zhou et al., IEEE/ASME Transactions on Mechatronics (2022), Fig. 3, p. 3",
   "source_url": "https://doi.org/10.1109/tmech.2022.3188955",
   "where": "Fig. 3, p. 3",
   "licence": "publisher copyright"
  },
  {
   "id": "cathpilot_zhou2022_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "CAD of the handle unit: shell, cam surface, sensor module, tensioner",
   "source_key": "zhou2022_cathpilot",
   "source": "Zhou et al., IEEE/ASME Transactions on Mechatronics (2022), Fig. 4, p. 4",
   "source_url": "https://doi.org/10.1109/tmech.2022.3188955",
   "where": "Fig. 4, p. 4",
   "licence": "publisher copyright"
  },
  {
   "id": "cathpilot_zhou2022_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Spring-loaded tensioner pulleys and brake surfaces inside the handle, top view",
   "source_key": "zhou2022_cathpilot",
   "source": "Zhou et al., IEEE/ASME Transactions on Mechatronics (2022), Fig. 6, p. 6",
   "source_url": "https://doi.org/10.1109/tmech.2022.3188955",
   "where": "Fig. 6, p. 6",
   "licence": "publisher copyright"
  },
  {
   "id": "cathpilot_zhou2022_06",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Force test layout: CathPilot, phantom, load cell and GUI",
   "source_key": "zhou2023_cathpilot",
   "source": "Zhou et al., IEEE Transactions on Biomedical Engineering (2023), Fig. 5, p. 5",
   "source_url": "https://doi.org/10.1109/tbme.2023.3276199",
   "where": "Fig. 5, p. 5",
   "licence": "publisher copyright"
  },
  {
   "id": "cathpilot_zhou2022_07",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Occlusion-crossing phantom setup with camera, GUI and delivery sheath",
   "source_key": "zhou2022_cathpilot",
   "source": "Zhou et al., IEEE/ASME Transactions on Mechatronics (2022), Fig. 10, p. 8",
   "source_url": "https://doi.org/10.1109/tmech.2022.3188955",
   "where": "Fig. 10, p. 8",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "zhou2022_cathpilot",
   "doi": "10.1109/tmech.2022.3188955",
   "url": null,
   "title": "The CathPilot: A Novel Approach for Accurate Interventional Device Steering and Tracking",
   "year": 2022,
   "venue": "IEEE/ASME Transactions on Mechatronics"
  },
  {
   "key": "zhou2023_cathpilot",
   "doi": "10.1109/tbme.2023.3276199",
   "url": null,
   "title": "The CathPilot: Performance Validation and Preclinical Safety and Feasibility Assessment",
   "year": 2023,
   "venue": "IEEE Transactions on Biomedical Engineering"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (TMECH 2022, TBME 2023)",
 "notes": "Not motorised: the 'teleoperation' is a cable-and-cam transmission; included as a boundary case because the coordinator's snowball list names it. Tavallaei is the author of the Western University RCNS papers (western_thakur2009).",
 "maintainers_own": false,
 "found_by": "gap"
}