 "schema_version": 1,
 "id": "cathbot_kundrat2021",
 "name": "CathBot",
 "full_name": "CathBot (MR-safe platform)",
 "group": "Imperial College London, Hamlyn Centre (Dagnino, Kundrat, Yang)",
 "group_short": "Imperial",
 "family": "academic",
 "category": "academic",
 "first_year": 2018,
 "latest_year": 2023,
 "versions": "2018 haptic active constraints (Dagnino IROS); 2021 MR-safe platform (Kundrat TBME); 2023 in-vivo platform (Dagnino TBME)",
 "target_anatomy": "renal, mesenteric, iliac and carotid arteries (pig); thoracic/abdominal phantoms",
 "devices": [
  "guidewire",
  "catheter"
 ],
 "devices_driven": "one catheter and one guidewire, one at a time (not simultaneously); standard instruments, plug-and-play carriers",
 "n_dof": 6,
 "n_dof_detail": "6 per Kundrat Sec. II-C (three actuated axes per instrument); 4 per LegoBot Table I",
 "translation_mechanism": "clamp-relay",
 "translation_detail": "clamp-relay (pneumatic clamps with pneumatic stepper drivers, regripping; single stroke 20 mm per review Table IV)",
 "rotation_mechanism": "other",
 "rotation_detail": "pneumatic rotary stepper (0.3 mm and 10 deg steps)",
 "device_range": "N.R. (supplement not opened); 6 Fr sheath in the pig study",
 "footprint_mm": "520 x 140 x 120 (follower); leader 191 x 111 x 85",
 "mass_kg": "N.R.",
 "force_sensing": "none on follower; haptics rendered from image-derived tip-to-wall distance",
 "haptics_leader": "yes: handle with linear and rotary motors renders viscous force/torque; virtual fixtures slow the follower near the wall",
 "imaging_compat": "MR-Safe (ASTM F2503 class); used under fluoroscopy",
 "sterility": "single-use concept; instrument carriers detachable",
 "highest_validation": "animal",
 "validation_detail": "4 pigs, 5 target arteries, cannulation and balloon angioplasty, robot vs manual; earlier phantom user study with 7 vascular surgeons/registrars",
 "key_numbers": "success 100% both; completion 35+-14 min robot vs 15+-6 min manual (p=0.03); fluoroscopy 7+-5 vs 2+-1 min; feed 7-10 mm/s, max 30 mm/s; actuator latency 10-15 ms; clamp holding force 10.2 N (calc.)",
 "sim_model": "none found for the robot (Kundrat/Dagnino texts)",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found (Hamlyn learning-from-demonstration studies listed by search, e.g. Chi ICRA 2020, were not matched to CathBot hardware)",
 "picture": {
  "status": "not-cleared",
  "grade": "A",
  "kind": "photo",
  "source": "abdelaziz2019_versatile Fig. 1, right part (MR-safe slave robot, studio photo; embedded image 3396x1600 on p1)",
  "source_url": "https://doi.org/10.1109/iros40897.2019.8968237",
  "licence": "publisher copyright (IEEE)",
  "credit": "Abdelaziz et al., 2019 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (2019), Fig. 1. publisher copyright (IEEE).",
  "edits": "native extract; cropped to the slave robot (master device on the left excluded); no erasing; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "cathbot_kundrat2021_01",
   "status": "cleared",
   "file": "pictures/extra/cathbot_kundrat2021_01.jpg",
   "thumb": "pictures/thumbs/cathbot_kundrat2021_01.jpg",
   "kind": "in-use",
   "caption": "In-vivo porcine trial: remote manipulator on mounting arm, master in control room",
   "source_key": "dagnino2023_in",
   "source": "Dagnino et al., IEEE Transactions on Biomedical Engineering (2023), Fig. 2(A)-(D), p. 2",
   "source_url": "https://doi.org/10.1109/tbme.2022.3227734",
   "where": "Fig. 2(A)-(D), p. 2",
   "licence": "CC BY 4.0"
  },
  {
   "id": "cathbot_kundrat2021_02",
   "status": "cleared",
   "file": "pictures/extra/cathbot_kundrat2021_02.jpg",
   "thumb": "pictures/thumbs/cathbot_kundrat2021_02.jpg",
   "kind": "in-use",
   "caption": "MR-safe slave robot driving devices into an aortic phantom under fluoroscopy",
   "source_key": "kundrat2021_mr",
   "source": "Kundrat et al., IEEE Transactions on Biomedical Engineering (2021), Fig. 4(b), p. 6",
   "source_url": "https://doi.org/10.1109/tbme.2021.3065146",
   "where": "Fig. 4(b), p. 6",
   "licence": "CC BY 4.0"
  },
  {
   "id": "cathbot_kundrat2021_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Slave robot on a patient phantom inside the MRI scanner bore",
   "source_key": "abdelaziz2019_versatile",
   "source": "Abdelaziz et al., 2019 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (2019), Fig. 6(a), p. 8",
   "source_url": "https://doi.org/10.1109/iros40897.2019.8968237",
   "where": "Fig. 6(a), p. 8",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "cathbot_kundrat2021_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "CAD close-up of pneumatic clamps and quick-release lid on the drive rails",
   "source_key": "abdelaziz2019_versatile",
   "source": "Abdelaziz et al., 2019 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (2019), Fig. 3(c), p. 3",
   "source_url": "https://doi.org/10.1109/iros40897.2019.8968237",
   "where": "Fig. 3(c), p. 3",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "cathbot_kundrat2021_05",
   "status": "cleared",
   "file": "pictures/extra/cathbot_kundrat2021_05.jpg",
   "thumb": "pictures/thumbs/cathbot_kundrat2021_05.jpg",
   "kind": "render",
   "caption": "Rendered platform: valve manifold, pneumatic tubing, master device and remote manipulator",
   "source_key": "dagnino2023_in",
   "source": "Dagnino et al., IEEE Transactions on Biomedical Engineering (2023), Fig. 1, p. 2",
   "source_url": "https://doi.org/10.1109/tbme.2022.3227734",
   "where": "Fig. 1, p. 2",
   "licence": "CC BY 4.0"
  },
  {
   "id": "cathbot_kundrat2021_06",
   "status": "cleared",
   "file": "pictures/extra/cathbot_kundrat2021_06.jpg",
   "thumb": "pictures/thumbs/cathbot_kundrat2021_06.jpg",
   "kind": "render",
   "caption": "CAD side view of catheter and guidewire drivers with quick-release lid details",
   "source_key": "kundrat2021_mr",
   "source": "Kundrat et al., IEEE Transactions on Biomedical Engineering (2021), Fig. 3(b), p. 4",
   "source_url": "https://doi.org/10.1109/tbme.2021.3065146",
   "where": "Fig. 3(b), p. 4",
   "licence": "CC BY 4.0"
  },
  {
   "id": "cathbot_kundrat2021_07",
   "status": "cleared",
   "file": "pictures/extra/cathbot_kundrat2021_07.jpg",
   "thumb": "pictures/thumbs/cathbot_kundrat2021_07.jpg",
   "kind": "leader",
   "caption": "Haptic master device with the navigation user interface in the control room",
   "source_key": "kundrat2021_mr",
   "source": "Kundrat et al., IEEE Transactions on Biomedical Engineering (2021), Fig. 4(a), p. 6",
   "source_url": "https://doi.org/10.1109/tbme.2021.3065146",
   "where": "Fig. 4(a), p. 6",
   "licence": "CC BY 4.0"
  },
  {
   "id": "cathbot_kundrat2021_08",
   "status": "cleared",
   "file": "pictures/extra/cathbot_kundrat2021_08.jpg",
   "thumb": "pictures/thumbs/cathbot_kundrat2021_08.jpg",
   "kind": "leader",
   "caption": "Master device prototype opened: belt drive, rotary and linear motors, force sensor",
   "source_key": "kundrat2021_mr",
   "source": "Kundrat et al., IEEE Transactions on Biomedical Engineering (2021), Fig. 2(b), p. 3",
   "source_url": "https://doi.org/10.1109/tbme.2021.3065146",
   "where": "Fig. 2(b), p. 3",
   "licence": "CC BY 4.0"
  }
 ],
 "sources": [
  {
   "key": "kundrat2021_mr",
   "doi": "10.1109/tbme.2021.3065146",
   "url": null,
   "title": "An MR-Safe Endovascular Robotic Platform: Design, Control, and Ex-Vivo Evaluation",
   "year": 2021,
   "venue": "IEEE Transactions on Biomedical Engineering"
  },
  {
   "key": "dagnino2023_in",
   "doi": "10.1109/tbme.2022.3227734",
   "url": null,
   "title": "In-Vivo Validation of a Novel Robotic Platform for Endovascular Intervention",
   "year": 2023,
   "venue": "IEEE Transactions on Biomedical Engineering"
  },
  {
   "key": "dagnino2018_haptic",
   "doi": "10.1109/iros.2018.8593628",
   "url": null,
   "title": "Haptic Feedback and Dynamic Active Constraints for Robot-Assisted Endovascular Catheterization",
   "year": 2018,
   "venue": "2018 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS)"
  },
  {
   "key": "pescio2025_endovascular",
   "doi": "10.1080/13645706.2025.2454237",
   "url": null,
   "title": "Endovascular robotics: technical advances and future directions",
   "year": 2025,
   "venue": "Minimally Invasive Therapy & Allied Technologies"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (kundrat2021_mr, dagnino2023_in); numbers also from",
 "notes": "LegoBot Table I lists ICL as 190x90x64 mm, clamp-based linear, 20 mm single stroke; Kundrat gives 520x140x120 mm. Row id keeps the proposed name.",
 "maintainers_own": false,
 "found_by": "academic"
}