 "schema_version": 1,
 "id": "casiattvr_lin2024",
 "name": "CASIA TTVR robot",
 "full_name": "CASIA robotic transcatheter tricuspid valve replacement (TTVR) system",
 "group": "Shuangyi Wang group, CASIA / University of Chinese Academy of Sciences and CAS Centre for Artificial Intelligence and Robotics, Hong Kong Institute of Science & Innovation, with Zeng-Guang Hou and Xiao-Hu Zhou; clinical partner Kent Chak-Yu So (CUHK, Prince of Wales Hospital)",
 "group_short": "CASIA",
 "family": "academic",
 "category": "academic",
 "first_year": 2024,
 "latest_year": 2026,
 "versions": "2024 TTVR system: passive stabilizer, robotic drive, detachable delivery catheter, valve manipulation mechanism; RL + Monte Carlo probabilistic maps + human-robot co-piloted control (wang2024_robotic, arXiv abstract); 2025 model-free delivery with Data-Enabled Predictive Control (DeePC) on the same drive and the LuX-Valve Plus delivery system, a camera standing in for fluoroscopy over a printed heart phantom (lin2025_modelfree, full); 2026 patient-specific geometry-aware planning + numerical-Jacobian IK control (lin2026_autonomous, abstract); 2026 virtual-physical simulation platform (lin2026_robotic, full)",
 "target_anatomy": "right heart: tricuspid valve (transcatheter tricuspid valve replacement / intervention)",
 "devices": [
  "catheter",
  "therapeutic-device"
 ],
 "devices_driven": "large-bore steerable TTVR delivery catheter (detachable) with valve Delivery system: LuX-Valve Plus (Jenscare Biotechnology, Ningbo) (lin2025 Sec. II).",
 "n_dof": 5,
 "n_dof_detail": "5 (overall forward/backward along the long axis, catheter rotation, outer-sheath advance/retract, bending of the bending tube, inner-core advance/retract; lin2025 Sec. II, Fig. 1(b)); 2 controlled (overall translation and bending; lin2025 Sec. III; lin2026_robotic Experimental setup)",
 "translation_mechanism": "other",
 "translation_detail": "other: robotic drive of the delivery-catheter system; overall translation by its own motor, drawn as a pinion on a toothed rack in lin2025 Fig. 1(b) (the text does not describe the mechanism)",
 "rotation_mechanism": "handle-rotation",
 "rotation_detail": "four motors in the drive unit, coupled through isolation parts to the instrument case that holds the delivery system (rotation, outer sheath, bending, inner core); \"The power from the Drive Unit is transmitted to the catheter through the Isolation Parts without direct physical contact\" (lin2025 Sec. II, Fig. 1(b))",
 "device_range": "workspace drawn in lin2025 Fig. 1(c): overall translation -28 to 212 mm, rotation -180 to 180 deg, outer sheath -107 to 20 mm, bending -90 to 0 deg, inner core -20 to 20 mm (graphic only, not in the text)",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "N.R.",
 "haptics_leader": "leader side with operator interface and GUI (lin2026_robotic Fig. 2b); haptics N.R.",
 "imaging_compat": "fluoroscopy (virtual fluoroscopy in the platform, 32 deg RAO / 7 deg CAU); optical shape/pose tracking of the catheter lin2025: a 640 x 480 px, 30 fps camera over the phantom stands in for fluoroscopy (Sec. IV-A).",
 "sterility": "components above the isolation parts can be sterilised during surgery (lin2025 Sec. II: \"the ability to sterilize the components above the Isolation Parts during surgery, while maintaining a sterile environment for the catheter\")",
 "highest_validation": "animal",
 "validation_detail": "first-in-vivo animal study (species and number N.R. in the arXiv abstract; full text not read); 15 patient CT datasets + 3 cardiac phantoms with mm-scale tip tracking error (lin2026_autonomous abstract); 10 operators raw vs augmented virtual fluoroscopy and 10 closed-loop control trials on the virtual-physical platform (lin2026_robotic) lin2025: printed patient-CT heart phantom (SVC, RA, RV); 3 catheters x 3 random start positions and 3 camera poses x 3 runs (Sec. IV-A, IV-C, Tables I-II); still phantom level.",
 "key_numbers": "co-piloted control better than master-slave in time efficiency, control efficiency, autonomy and stability (numbers N.R. in abstract); platform: tracking-to-CT registration 0.59 +/- 0.16 mm, landmark projection 2.31 +/- 0.60 px, input-to-display latency 255.6 +/- 13.4 ms, 2D angular error 15.02 -> 2.57 deg (median) with augmented fluoroscopy across 10 operators (lin2026_robotic Tables 1-2) lin2025 DeePC: \"In most cases, the final error in slope is less than 1 deg and the error in intercept is under 3 pixels\" for the tip line in the camera image (Sec. IV-C); worst case 2.55 +/- 0.04 deg and 3.37 +/- 0.17 px (camera pose 3, position 2, Table II); no success rate or time reported.",
 "sim_model": "virtual-physical simulator: real robot-driven catheter, optical shape tracking, CT-derived virtual fluoroscopy in 3D Slicer 5.10 (lin2026_robotic)",
 "has_sim_model": true,
 "sim2real": "unclear",
 "sim2real_detail": "RL is part of the 2024 system (wang2024_robotic abstract); where it was trained is N.R. (arXiv PDF not read). lin2025_modelfree (full): DeePC uses a Hankel matrix of the operator's recorded master-slave motion and \"does not require additional calibration or model training\" (Sec. III-A, V); no simulator.",
 "picture": {
  "status": "not-cleared",
  "grade": "A",
  "kind": "photo",
  "source": "lin2026_autonomous Fig. 6(a) (experimental set-up, p. 9), IEEE TBME 2026,",
  "source_url": "https://doi.org/10.1109/tbme.2026.3725663",
  "licence": "publisher copyright (IEEE)",
  "credit": "Lin et al., IEEE Transactions on Biomedical Engineering (2026), Fig. 6(a). publisher copyright (IEEE).",
  "edits": "native extract: the photo is embedded as three 3302 px raster strips (p9_9, p9_7, p9_12; the labels and arrows are vector text, not in the raster); stitched without overlap to 3302 x 689; cropped to the robot (left 44.5 %, phantom and holders cut); nothing erased; none upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "casiattvr_lin2024_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Driving system with delivery catheter, camera measuring system and heart phantom",
   "source_key": "lin2025_modelfree",
   "source": "Lin et al., 2025 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (2025), Fig. 5, p. 4",
   "source_url": "https://doi.org/10.1109/iros60139.2025.11247226",
   "where": "Fig. 5, p. 4",
   "licence": "publisher copyright"
  },
  {
   "id": "casiattvr_lin2024_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Drive unit: instrument case, motors, isolation parts and controller",
   "source_key": "lin2025_modelfree",
   "source": "Lin et al., 2025 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (2025), Fig. 1(b), p. 2",
   "source_url": "https://doi.org/10.1109/iros60139.2025.11247226",
   "where": "Fig. 1(b), p. 2",
   "licence": "publisher copyright"
  },
  {
   "id": "casiattvr_lin2024_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Robotic drive holding the LuX-Valve delivery catheter",
   "source_key": "lin2025_modelfree",
   "source": "Lin et al., 2025 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (2025), Fig. 1(a), p. 2",
   "source_url": "https://doi.org/10.1109/iros60139.2025.11247226",
   "where": "Fig. 1(a), p. 2",
   "licence": "publisher copyright"
  },
  {
   "id": "casiattvr_lin2024_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Delivery catheter in the instrument case with drive, valve-rotation and translation modules",
   "source_key": "lin2026_autonomous",
   "source": "Lin et al., IEEE Transactions on Biomedical Engineering (2026), Fig. 2, p. 6",
   "source_url": "https://doi.org/10.1109/tbme.2026.3725663",
   "where": "Fig. 2, p. 6",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "wang2024_robotic",
   "doi": "10.48550/ARXIV.2411.12478",
   "url": null,
   "title": "Robotic transcatheter tricuspid valve replacement with hybrid enhanced intelligence: a new paradigm and first-in-vivo study",
   "year": 2024,
   "venue": "arXiv"
  },
  {
   "key": "lin2025_modelfree",
   "doi": "10.1109/iros60139.2025.11247226",
   "url": null,
   "title": "Model-Free Catheter Delivery Strategy for Robotic Transcatheter Tricuspid Valve Replacement",
   "year": 2025,
   "venue": "2025 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS)"
  },
  {
   "key": "lin2026_autonomous",
   "doi": "10.1109/tbme.2026.3725663",
   "url": null,
   "title": "Autonomous Patient-Specific Geometry-Aware Planning and Control Framework for Robotic Transcatheter Tricuspid Valve Intervention",
   "year": 2026,
   "venue": "IEEE Transactions on Biomedical Engineering"
  },
  {
   "key": "lin2026_robotic",
   "doi": "10.1007/s11701-026-03852-1",
   "url": null,
   "title": "A robotic simulation platform for fluoroscopy-guided catheter navigation in transcatheter tricuspid valve intervention: system development and proof-of-concept applications",
   "year": 2026,
   "venue": "Journal of Robotic Surgery"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (lin2026_robotic); abstract (wang2024_robotic, lin2026_autonomous); full (lin2025_modelfree, 2026-10-14)",
 "notes": "Structural-heart catheter robot, not a guidewire/neuro robot; include or move to a structural-heart box.",
 "maintainers_own": false,
 "found_by": "gap"
}