 "schema_version": 1,
 "id": "beihang_wang2010",
 "name": "Beihang VIR",
 "full_name": "Beihang remote-controlled vascular interventional robot (VISR / VIR / RVIR / VIR-2) and its successor catheter operating robot (COR, 2016), the 2023 MIVI robot",
 "group": "Beihang University Robotics Institute (Tianmiao Wang, Da Liu; 2010 corresponding author Dapeng Zhang) and Image Processing Centre / School of Astronautics (Cai Meng); clinical: Department of Neurosurgery, Navy General Hospital of PLA (Wang-sheng Lu, Zeng-min Tian; author affiliation in lu2011 and lu2016); 2023 version Beijing Institute of Petrochemical Technology (Shaoya Guan, Limei Ma) with Cai Meng",
 "group_short": "Beihang U.",
 "family": "commercial",
 "category": "clinical-stage",
 "first_year": 2010,
 "latest_year": 2023,
 "versions": "2010 VISR: 5-DOF lockable passive hydraulic supporting arm + catheter navigator (rollers on a gear), PVDF tip sensor, Omega 6 master, glass model and 1 dog (wang2010_remote) -> 2011 VIR + virtual surgery training system, 10 beagles (lu2011_application) -> 2013 RVIR: image guidance (C-arm calibration, MicronTracker, catheter localisation, 3D reconstruction), Falcon master (meng2013_remote) -> VIR-2, 15 patients 2013 (lu2016_clinical, online 2015) -> 2016 COR, a new driver with bionic friction fingers, leadscrew rotation and a clamping motor, 2.6 kg, bench and glass model (meng2016_novel) -> 2023 MIVI robot with EM-tracked navigation (guan2023_multi; its Fig. 5(b) is the same photograph as meng2016 Fig. 5)",
 "target_anatomy": "cerebral (angiography in patients); renal, cardiac chambers, cerebrovascular arteries in animals; aortic (2023 navigation: AAA, TAA, AD data)",
 "devices": [
  "guidewire",
  "catheter",
  "sheath"
 ],
 "devices_driven": "catheter only, one device: 1.5 mm catheters (wang2010_remote p. 200); 4F-6F (1.33-2.00 mm) single-bend catheters through a 6F sheath, guidewire not driven ('the medical robot controlled only the catheter', lu2011_application p. 363, 365); 5F single-curve catheter in patients (lu2016_clinical p. 133); COR: 4F catheter (meng2016_novel p. 307)",
 "n_dof": 2,
 "n_dof_detail": "2 active (axial: stepping motor 2; rotation: stepping motor 1) on a 5-DOF lockable passive hydraulic supporting arm (wang2010_remote p. 196); COR: translation, rotation and clamping motors (meng2016_novel p. 305)",
 "translation_mechanism": "friction-roller",
 "translation_detail": "friction-roller (catheter clamped by rollers 1 and 2 driven by stepping motor 2, wang2010_remote p. 196; COR / 2023 MIVI: bionic friction fingers, meng2016_novel p. 305, guan2023_multi Fig. 5a)",
 "rotation_mechanism": "other",
 "rotation_detail": "the roller pair sits on a passive gear turned by stepping motor 1 through an active gear (wang2010_remote p. 196); COR: the two bionic fingers translate in opposite directions via a two-direction leadscrew and belt (finger rolling) (meng2016_novel p. 305); 2023: rotation motor (guan2023_multi Fig. 5a)",
 "device_range": "catheter speed 10-25 mm/s (wang2010_remote p. 199; lu2011_application p. 364); stroke N.R.; COR 205 pulses per mm (meng2016_novel p. 307)",
 "footprint_mm": "N.R.",
 "mass_kg": "COR 2.6 kg, aluminium alloy and steel (meng2016_novel p. 309); VIR N.R.",
 "force_sensing": "PVDF piezoelectric film glued at the catheter tip, robot stalls above a threshold (wang2010_remote p. 198); lu2011 calls the force feedback virtual (p. 365); not used in the 15 patients (lu2016_clinical p. 136); COR none",
 "haptics_leader": "Omega 6 haptic device (Force Dimension) (wang2010_remote p. 198); 3-DOF Novint Falcon in a shielded cabinet (meng2013_remote p. 231); master controlled over LAN, device N.R. (lu2016_clinical p. 133)",
 "imaging_compat": "X-ray/DSA (GE LCV+ C-arm); RVIR image guidance with an H40 MicronTracker and C-arm calibration phantom (meng2013_remote p. 231-236); 2023: EM tracking + CTA-DSA 2D/3D registration",
 "sterility": "N.R.",
 "highest_validation": "human",
 "validation_detail": "15 patients (March-September 2013), robot-assisted cerebral angiography, femoral sheath placed by hand, no complications (lu2016_clinical p. 133-134); 10 beagles (16-20 kg), 9 targets including both carotid and both vertebral arteries, 5 repeats, no complications (lu2011_application p. 363; meng2013_remote '10 dogs'); 1 dog, 5 targets, and glass model with 10 repeats per speed (wang2010_remote); COR: bench (10 trials per length 10-50 mm, 10 rotations) and glass model (meng2016_novel); 2023 vascular model only (guan2023_multi)",
 "key_numbers": "remote positioning accuracy 1.05 +/- 0.28 mm; staff DSA exposure 0 min; procedure 25-41 min, mean 34.4 +/- 5.13 min (lu2016_clinical p. 134); catheter motion error < 0.5 mm in a glass model (wang2010_remote p. 199); distortion correction RMS 0.35 px, distance reconstruction RMS 0.53 mm (max 1.34 mm) (meng2013_remote); COR axial accuracy -0.04 to 0.34 mm, relative error within +-5% (meng2016_novel Table I); 2023: 2D/3D registration < 1 mm, EM positioning RMS < 1 mm and max < 2 mm over 4 paths x 26 points (guan2023_multi Results, Fig. 7)",
 "sim_model": "virtual surgery training system (lu2011_application abstract); no controller-training model",
 "has_sim_model": true,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "B",
  "kind": "photo",
  "source": "lu2016_clinical Fig. 2 (slave component of the propulsion system of the VIR-2 robot used in the 15-patient series), Int J Med Robot 2016;12:132-136, p. 133,",
  "source_url": "https://doi.org/10.1002/rcs.1650",
  "licence": "publisher copyright (Wiley)",
  "credit": "Lu et al., The International Journal of Medical Robotics and Computer Assisted Surgery (2015), Fig. 2. publisher copyright (Wiley).",
  "edits": "native extract (953 x 693); 3-px white frame cropped; nothing erased; none upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "beihang_wang2010_01",
   "status": "cleared",
   "file": "pictures/extra/beihang_wang2010_01.jpg",
   "thumb": "pictures/thumbs/beihang_wang2010_01.jpg",
   "kind": "render",
   "caption": "CAD of the MIVI robot: bionic fingers, clamping and rotation motors",
   "source_key": "guan2023_multi",
   "source": "Guan et al., BMC Surgery (2023), Fig. 5(a), p. 4",
   "source_url": "https://doi.org/10.1186/s12893-023-01944-5",
   "where": "Fig. 5(a), p. 4",
   "licence": "CC BY 4.0"
  },
  {
   "id": "beihang_wang2010_02",
   "status": "cleared",
   "file": "pictures/extra/beihang_wang2010_02.jpg",
   "thumb": "pictures/thumbs/beihang_wang2010_02.jpg",
   "kind": "in-use",
   "caption": "Catheter navigated to four target positions in a bifurcation phantom",
   "source_key": "guan2023_multi",
   "source": "Guan et al., BMC Surgery (2023), Fig. 6, p. 4",
   "source_url": "https://doi.org/10.1186/s12893-023-01944-5",
   "where": "Fig. 6, p. 4",
   "licence": "CC BY 4.0"
  },
  {
   "id": "beihang_wang2010_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Supporting manipulator and catheter navigator at a glass vascular model under the C-arm",
   "source_key": "meng2013_remote",
   "source": "Meng et al., The International Journal of Medical Robotics and Computer Assisted Surgery (2013), Fig. 12, p. 238",
   "source_url": "https://doi.org/10.1002/rcs.1494",
   "where": "Fig. 12, p. 238",
   "licence": "publisher copyright"
  },
  {
   "id": "beihang_wang2010_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Robot beside a dog on the angiography table during the animal experiment",
   "source_key": "meng2013_remote",
   "source": "Meng et al., The International Journal of Medical Robotics and Computer Assisted Surgery (2013), Fig. 13, p. 238",
   "source_url": "https://doi.org/10.1002/rcs.1494",
   "where": "Fig. 13, p. 238",
   "licence": "publisher copyright"
  },
  {
   "id": "beihang_wang2010_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Clinical cerebral angiography: catheter fixed in the slave, bedside scene, operator at the master",
   "source_key": "lu2016_clinical",
   "source": "Lu et al., The International Journal of Medical Robotics and Computer Assisted Surgery (2015), Fig. 3(a-d), p. 134",
   "source_url": "https://doi.org/10.1002/rcs.1650",
   "where": "Fig. 3(a-d), p. 134",
   "licence": "publisher copyright"
  },
  {
   "id": "beihang_wang2010_06",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "RVIR master site (image guidance, remote control device) and slave site (labelled)",
   "source_key": "meng2013_remote",
   "source": "Meng et al., The International Journal of Medical Robotics and Computer Assisted Surgery (2013), Fig. 1, p. 232",
   "source_url": "https://doi.org/10.1002/rcs.1494",
   "where": "Fig. 1, p. 232",
   "licence": "publisher copyright"
  },
  {
   "id": "beihang_wang2010_07",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "photo",
   "caption": "2016 catheter operating robot (COR): prototype photo and CAD with bionic fingers",
   "source_key": "meng2016_novel",
   "source": "Meng et al., 2016 IEEE Workshop on Advanced Robotics and its Social Impacts (ARSO) (2016), Fig. 1(a-b), p. 305",
   "source_url": "https://doi.org/10.1109/arso.2016.7736300",
   "where": "Fig. 1(a-b), p. 305",
   "licence": "publisher copyright"
  },
  {
   "id": "beihang_wang2010_08",
   "status": "cleared",
   "file": "pictures/extra/beihang_wang2010_08.jpg",
   "thumb": "pictures/thumbs/beihang_wang2010_08.jpg",
   "kind": "close-up",
   "caption": "Bionic-finger catheter driver (COR / MIVI robot) during a rotation test",
   "source_key": "guan2023_multi",
   "source": "Guan et al., BMC Surgery (2023), Fig. 5(b), p. 4 (previous main picture; the same photo is meng2016_novel Fig. 5)",
   "source_url": "https://doi.org/10.1186/s12893-023-01944-5",
   "where": "Fig. 5(b), p. 4 (previous main picture; the same photo is meng2016_novel Fig. 5)",
   "licence": "CC BY 4.0"
  }
 ],
 "sources": [
  {
   "key": "wang2010_remote",
   "doi": "10.1002/rcs.306",
   "url": null,
   "title": "Remote‐controlled vascular interventional surgery robot",
   "year": 2010,
   "venue": "The International Journal of Medical Robotics and Computer Assisted Surgery"
  },
  {
   "key": "lu2011_application",
   "doi": "10.1002/rcs.406",
   "url": null,
   "title": "Application study of medical robots in vascular intervention",
   "year": 2011,
   "venue": "The International Journal of Medical Robotics and Computer Assisted Surgery"
  },
  {
   "key": "meng2013_remote",
   "doi": "10.1002/rcs.1494",
   "url": null,
   "title": "A remote‐controlled vascular interventional robot: system structure and image guidance",
   "year": 2013,
   "venue": "The International Journal of Medical Robotics and Computer Assisted Surgery"
  },
  {
   "key": "lu2016_clinical",
   "doi": "10.1002/rcs.1650",
   "url": null,
   "title": "Clinical application of a vascular interventional robot in cerebral angiography",
   "year": 2015,
   "venue": "The International Journal of Medical Robotics and Computer Assisted Surgery"
  },
  {
   "key": "meng2016_novel",
   "doi": "10.1109/arso.2016.7736300",
   "url": null,
   "title": "A novel catheter operating robot for vascular interventional surgery",
   "year": 2016,
   "venue": "2016 IEEE Workshop on Advanced Robotics and its Social Impacts (ARSO)"
  },
  {
   "key": "guan2023_multi",
   "doi": "10.1186/s12893-023-01944-5",
   "url": null,
   "title": "Multi-mode information fusion navigation system for robot-assisted vascular interventional surgery",
   "year": 2023,
   "venue": "BMC Surgery"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (wang2010_remote, lu2011_application, meng2013_remote, lu2016_clinical, meng2016_novel, guan2023_multi)",
 "notes": "Two hardware lines in one row: VISR/VIR/RVIR/VIR-2 (2010-2016, human) and the COR (2016) = 2023 MIVI robot (phantom). guan2023 Fig. 5(b) and meng2016 Fig. 5 are the same photograph (compared side by side by X2), so the MIVI robot is the COR; meng2016 calls the COR a new robot and cites wang2010 only as related work. | X2 2026-11-08: all five Beihang papers read in full; main picture now the VIR-2 drive (lu2016 Fig. 2).",
 "maintainers_own": false,
 "found_by": "gap"
}