 "schema_version": 1,
 "id": "armm_heunis2020",
 "name": "ARMM",
 "full_name": "ARMM (Advanced Robotics for Magnetic Manipulation)",
 "group": "Misra, Surgical Robotics Lab, University of Twente",
 "group_short": "U. Twente",
 "family": "magnetic",
 "category": "magnetic-navigation",
 "first_year": 2019,
 "latest_year": 2020,
 "versions": "single (RA-L 2020); coil optimisation in Sikorski 2019 (cited, not read); 2016: Twente (Misra) magnetically actuated catheter in closed loop on 2D ultrasound (Boskma, Scheggi, Misra 2016 BioRob) (Alekseeva 2025 Table 12 row 6) (alekseeva2025_design)",
 "target_anatomy": "iliac / aortic phantom from CT of a 62-year-old (gelatin torso phantom, pulsating flow)",
 "devices": [
  "catheter"
 ],
 "devices_driven": "commercial Sherpa NX Active 2 mm coronary guide catheter (500 mm) with five NdFeB magnets (2 mm x 2 mm) in the tip",
 "n_dof": null,
 "n_dof_detail": "two UR manipulators (UR10 carries the electromagnetic coil, UR5 carries the ultrasound probe) plus a 1-axis catheter insertion device",
 "translation_mechanism": "magnetic",
 "translation_detail": "magnetic (steering); separate insertion device, servo-driven gear and bearing, 1.3-6.4 mm/s, takes 3-34 Fr",
 "rotation_mechanism": "magnetic-field",
 "rotation_detail": "coil pose, no shaft rotation",
 "device_range": "500 mm catheter",
 "footprint_mm": "workspace sphere radius 783 mm (max 1300 mm); 9 m3 environment",
 "mass_kg": "N.R.",
 "force_sensing": "none; tip tracked by ultrasound (30 Hz) and motion capture",
 "haptics_leader": "no",
 "imaging_compat": "3D ultrasound feedback, optical motion capture; radiation-free",
 "sterility": "N.R.",
 "highest_validation": "phantom",
 "validation_detail": "gelatin torso phantom with pulsating flow, 3D targets, PI control; target count N.R. in my read",
 "key_numbers": "coil at least 20 mT and over 0.6 mT/mm at 50-200 mm; mean tip error 2.09 +/- 0.49 mm; median time to target 32.6 s",
 "sim_model": "none found (dipole-field actuation model; no simulator)",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "A",
  "kind": "photo",
  "source": "U-Today 2021-06-28 \"Robot for surgeries inside blood vessels\", photo by Hans Wolkers: https://www.utoday.nl/uploads/images/2021/week26/_Z723444.jpg (UR arm with electromagnet over the torso phantom, researcher operating)",
  "source_url": "https://www.utoday.nl/uploads/images/2021/week26/_Z723444.jpg",
  "licence": "press image (U-Today / Hans Wolkers), terms not stated",
  "credit": "U-Today 2021-06-28 \"Robot for surgeries inside blood vessels\", photo by Hans Wolkers: https://www.utoday.nl/uploads/images/2021/week26/_Z723444.jpg (UR arm with electromagnet over the torso phantom, researcher operating). Licence: press image (U-Today / Hans Wolkers), terms not stated.",
  "edits": "downscaled 50 % from 5315 px (no crop); not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "armm_heunis2020_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Electromagnet coil with cooling fans on the UR arm end-effector",
   "source_key": "heunis2020_armm",
   "source": "Heunis et al., IEEE Robotics and Automation Letters (2020), U-Today 2021-06-28 photo by Hans Wolkers, (cropped to the coil)",
   "source_url": "https://doi.org/10.1109/lra.2020.2965077",
   "where": "U-Today 2021-06-28 photo by Hans Wolkers, https://www.utoday.nl/uploads/images/2021/week26/_Z723495.jpg (cropped to the coil)",
   "licence": "press image (U-Today / Hans Wolkers)"
  },
  {
   "id": "armm_heunis2020_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "ARMM in a clinical-style setting: UR arms with coil and US probe over a torso",
   "source_key": "heunis2020_armm",
   "source": "Heunis et al., IEEE Robotics and Automation Letters (2020), Fig. 1, p. 1",
   "source_url": "https://doi.org/10.1109/lra.2020.2965077",
   "where": "Fig. 1, p. 1",
   "licence": "publisher copyright"
  },
  {
   "id": "armm_heunis2020_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Rendering of the electromagnetic coil and the force-sensed ultrasound transducer over the phantom",
   "source_key": "heunis2020_armm",
   "source": "Heunis et al., IEEE Robotics and Automation Letters (2020), Fig. 4 (right), p. 5",
   "source_url": "https://doi.org/10.1109/lra.2020.2965077",
   "where": "Fig. 4 (right), p. 5",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "heunis2020_armm",
   "doi": "10.1109/lra.2020.2965077",
   "url": null,
   "title": "The ARMM System - Autonomous Steering of Magnetically-Actuated Catheters: Towards Endovascular Applications",
   "year": 2020,
   "venue": "IEEE Robotics and Automation Letters"
  },
  {
   "key": "alekseeva2025_design",
   "doi": "10.1002/rob.22480",
   "url": null,
   "title": "The Design of Endovascular Robots for Cardiac Interventions: A Survey",
   "year": 2024,
   "venue": "Journal of Field Robotics"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full",
 "notes": "Mobile electromagnet on a robot arm; not a clinical room install.",
 "maintainers_own": false,
 "found_by": "actuated-tip"
}