 "schema_version": 1,
 "id": "allvas_aopeng2021",
 "name": "ALLVAS",
 "full_name": "ALLVAS (REIS)",
 "group": "Aopeng Medical with Shanghai Changhai Hospital (Qingsheng Lu) and Shanghai Jiao Tong University (Kundong Wang) (song2022_biomimetic affiliations; company not named there)",
 "group_short": "Aopeng Medical",
 "family": "commercial",
 "category": "clinical-stage",
 "first_year": 2021,
 "latest_year": 2026,
 "versions": "ALLVAS; predecessor: the group's first-generation robot ('similar to the first-generation surgical robot initially studied by our research group', song2022_biomimetic p. 14, refs 13-14 = Wang 2018 IJMRCAS 14:e1915 and Lu 2020 VES 54:598-604; not read) -> 2022 biomimetic system, 38 porcine cases (song2022_biomimetic; robot unnamed there, identified as ALLVAS by song2023_novel, which cites it) -> 2023 EVAR in patients (song2023_novel); 2022: Changhai (Q. Lu) remote-controlled robot driven by hollow ultrasonic motors, ex vivo phantom (Lu 2022) (Alekseeva 2025 Table 7 row 1, Fig. 13e; Belikov 2024 Fig. 6 QIN) (alekseeva2025_design; belikov2024_evolution)",
 "target_anatomy": "aortic (EVAR), peripheral, coronary (one remote PCI)",
 "devices": [
  "guidewire",
  "catheter",
  "therapeutic-device"
 ],
 "devices_driven": "four manipulators with double V-grippers handle devices from a 0.014-in (0.36 mm) guidewire to a 40 mm stent-graft system (song2022_biomimetic p. 5); 2-3 devices at once in the porcine cases (catheter + guidewire, or stent-graft system, inner core and guidewire in TEVAR) (p. 10); up to 4 devices (song2023_novel); compatible with market devices",
 "n_dof": 18,
 "n_dof_detail": "18 (slave: two three-axis arms and four manipulators, each manipulator with two DC motors and a stepper; song2022_biomimetic p. 4, 7); '12-DOF motion in the four directions of the short axis' (p. 14); 12 per mandhumula2025_robotic as REIS",
 "translation_mechanism": "clamp-relay",
 "translation_detail": "clamp-relay",
 "rotation_mechanism": "rotating-clamp",
 "rotation_detail": "rotatable V-grippers, continuous progressive rotation with relay gripper (song2023_novel)",
 "device_range": "max axial speed 275 mm/s, min 5.15 mm/s, max thrust 27.45 N, max displacement 140 cm (song2022_biomimetic Table 1); rotation speeds 1.57 and 0.52 rad/s (p. 7)",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "force feedback; force sensor on the manipulator (type N.R.), in vitro feedback delay < 60 ms, peak error 0.15 N (song2022_biomimetic p. 8); automatic mode halts above 2 N (liang2025_unleashing)",
 "haptics_leader": "joysticks on a console outside the cath lab: button = clamp/release, knob = rotate, swing = advance/retract; force-feedback handle based on electromagnetic resistance (song2022_biomimetic p. 3, 7, 14; the paper says both 'four joysticks' and 'two sets of master joysticks')",
 "imaging_compat": "mobile base, arm + manipulator",
 "sterility": "detachable grippers, aimed at replacing all contact parts (song2022_biomimetic abstract, p. 14); single-use status N.R.",
 "highest_validation": "human",
 "validation_detail": "EVAR 5 pts (song2023_novel); 4 pts programmed steps (liang2025_unleashing); 1 remote PCI 800 km (Xu 2024); animal: 10 domestic swine (45 kg), 38 robotic cases (PCI 8, TEVAR 4, renal stent 8, iliac stent 10, IVC filter 8), 5 surgeons, plus a 3D-printed aorto-iliac model (song2022_biomimetic p. 9-12)",
 "key_numbers": "EVAR 5/5, operator dose -96.5% (song2023_novel abstract); remote PCI latency 30 ms, 800 km (chen2026_teleoperation Table I); swine: all technically successful, no adverse events, PCI operator dose 24.6 +- 2.72 uGy robotic vs 2224 +- 92.5 manual (the manual operator values equal the manual patient values in every column, possibly a table error), operative time not different, device exchange within 10 s, rotation accuracy 1 degree (song2022_biomimetic Table 2, p. 14)",
 "sim_model": "yes for vessels/wire: SOFA-based training platform of Tongji group (yao2025_sim4endor, full); no model of the robot itself found",
 "has_sim_model": true,
 "sim2real": "yes",
 "sim2real_detail": "YES: chen2026_learning (SAC + MPC filter, sim 49/50 -> phantom 18/20, full, IROS 2026); tian2026_visionlanguage (phantom 30/30, full); Tongji group lists 7 papers on ALLVAS/SOFA (S7b.md); liu2024_autonomous trained on robot not sim",
 "picture": {
  "status": "cleared",
  "grade": "C",
  "kind": "photo",
  "source": "tian2026_visionlanguage (arXiv 2606.30698) Fig. 3(a), native image xref 313 with the authors' soft mask, p5",
  "source_url": "https://arxiv.org/abs/2606.30698",
  "licence": "CC BY 4.0 (both arXiv licences checked on arxiv.org/abs)",
  "credit": "Tian et al., 2026, Fig. 3(a). CC BY 4.0 (both arXiv licences checked on arxiv.org/abs).",
  "edits": "full native image 777x690 with the authors' alpha, flattened on white; cut = authors' alpha (no SAM)",
  "file": "pictures/allvas_aopeng2021.jpg"
 },
 "pictures": [
  {
   "id": "allvas_aopeng2021_01",
   "status": "cleared",
   "file": "pictures/extra/allvas_aopeng2021_01.jpg",
   "thumb": "pictures/thumbs/allvas_aopeng2021_01.jpg",
   "kind": "close-up",
   "caption": "Draped mechanical arms carrying the rotatable gripper relays",
   "source_key": "chen2026_learning",
   "source": "Chen et al. (2026), Fig. 2, arXiv p. 3",
   "source_url": "https://arxiv.org/abs/2607.00066",
   "where": "Fig. 2, arXiv p. 3",
   "licence": "CC BY 4.0"
  },
  {
   "id": "allvas_aopeng2021_02",
   "status": "cleared",
   "file": "pictures/extra/allvas_aopeng2021_02.jpg",
   "thumb": "pictures/thumbs/allvas_aopeng2021_02.jpg",
   "kind": "render",
   "caption": "Functional unit: renderings and a photograph beside the angiography table",
   "source_key": "song2023_novel",
   "source": "Song et al., European Radiology (2023), Fig. 1, p. 3",
   "source_url": "https://doi.org/10.1007/s00330-023-09810-x",
   "where": "Fig. 1, p. 3",
   "licence": "CC BY 4.0"
  },
  {
   "id": "allvas_aopeng2021_03",
   "status": "cleared",
   "file": "pictures/extra/allvas_aopeng2021_03.jpg",
   "thumb": "pictures/thumbs/allvas_aopeng2021_03.jpg",
   "kind": "leader",
   "caption": "Rendering of the interventional console with monitors and control panels",
   "source_key": "song2023_novel",
   "source": "Song et al., European Radiology (2023), Fig. 2, p. 3",
   "source_url": "https://doi.org/10.1007/s00330-023-09810-x",
   "where": "Fig. 2, p. 3",
   "licence": "CC BY 4.0"
  },
  {
   "id": "allvas_aopeng2021_04",
   "status": "cleared",
   "file": "pictures/extra/allvas_aopeng2021_04.jpg",
   "thumb": "pictures/thumbs/allvas_aopeng2021_04.jpg",
   "kind": "close-up",
   "caption": "Renderings of the grippers rotating through four motion states",
   "source_key": "song2023_novel",
   "source": "Song et al., European Radiology (2023), Fig. 5, p. 8",
   "source_url": "https://doi.org/10.1007/s00330-023-09810-x",
   "where": "Fig. 5, p. 8",
   "licence": "CC BY 4.0"
  },
  {
   "id": "allvas_aopeng2021_05",
   "status": "cleared",
   "file": "pictures/extra/allvas_aopeng2021_05.jpg",
   "thumb": "pictures/thumbs/allvas_aopeng2021_05.jpg",
   "kind": "render",
   "caption": "Rendering of the two mechanical arms with gripper ends",
   "source_key": "tian2026_visionlanguage",
   "source": "Tian et al. (2026), Fig. 3(b), arXiv p. 5",
   "source_url": "https://arxiv.org/abs/2606.30698",
   "where": "Fig. 3(b), arXiv p. 5",
   "licence": "CC BY 4.0"
  },
  {
   "id": "allvas_aopeng2021_06",
   "status": "cleared",
   "file": "pictures/extra/allvas_aopeng2021_06.jpg",
   "thumb": "pictures/thumbs/allvas_aopeng2021_06.jpg",
   "kind": "render",
   "caption": "Bionic design: two three-axis arms with four manipulators and grippers (labelled)",
   "source_key": "song2022_biomimetic",
   "source": "Song et al., Micromachines (2022), Fig. 2(a-c), p. 4",
   "source_url": "https://doi.org/10.3390/mi13101587",
   "where": "Fig. 2(a-c), p. 4",
   "licence": "CC BY 4.0"
  },
  {
   "id": "allvas_aopeng2021_07",
   "status": "cleared",
   "file": "pictures/extra/allvas_aopeng2021_07.jpg",
   "thumb": "pictures/thumbs/allvas_aopeng2021_07.jpg",
   "kind": "close-up",
   "caption": "Rendering of the coaxial manipulators clamping a stent-graft system",
   "source_key": "song2022_biomimetic",
   "source": "Song et al., Micromachines (2022), Fig. 3(b), p. 6",
   "source_url": "https://doi.org/10.3390/mi13101587",
   "where": "Fig. 3(b), p. 6",
   "licence": "CC BY 4.0"
  },
  {
   "id": "allvas_aopeng2021_08",
   "status": "cleared",
   "file": "pictures/extra/allvas_aopeng2021_08.jpg",
   "thumb": "pictures/thumbs/allvas_aopeng2021_08.jpg",
   "kind": "in-use",
   "caption": "Robot grippers at the table beside the angiograms of five porcine procedure types",
   "source_key": "song2022_biomimetic",
   "source": "Song et al., Micromachines (2022), Fig. 8(a-e), p. 13",
   "source_url": "https://doi.org/10.3390/mi13101587",
   "where": "Fig. 8(a-e), p. 13",
   "licence": "CC BY 4.0"
  }
 ],
 "sources": [
  {
   "key": "song2023_novel",
   "doi": "10.1007/s00330-023-09810-x",
   "url": null,
   "title": "A novel endovascular robotic-assisted system for endovascular aortic repair: first-in-human evaluation of practicability and safety",
   "year": 2023,
   "venue": "European Radiology"
  },
  {
   "key": "song2022_biomimetic",
   "doi": "10.3390/mi13101587",
   "url": null,
   "title": "Novel Endovascular Interventional Surgical Robotic System Based on Biomimetic Manipulation",
   "year": 2022,
   "venue": "Micromachines"
  },
  {
   "key": "liang2025_unleashing",
   "doi": "10.1002/mco2.70489",
   "url": null,
   "title": "Unleashing the Potential: First-in-Human Evaluation of Automatic Robotic-Assisted Endovascular Aortic Repair for Standardized Therapies",
   "year": 2025,
   "venue": "MedComm"
  },
  {
   "key": "chen2026_teleoperation",
   "doi": "10.1109/TMRB.2026.3681030",
   "url": null,
   "title": "Remote Teleoperation of Endovascular Intervention Robots: A Systematic Review",
   "year": 2026,
   "venue": "IEEE Transactions on Medical Robotics and Bionics"
  },
  {
   "key": "chen2026_learning",
   "doi": null,
   "url": "https://arxiv.org/abs/2607.00066",
   "title": "Learning Expert Strategy for Autonomous Robotic Endovascular Intervention via Decoupled Procedural Execution",
   "year": 2026,
   "venue": null
  },
  {
   "key": "tian2026_visionlanguage",
   "doi": null,
   "url": "https://arxiv.org/abs/2606.30698",
   "title": "Vision-Language Procedural Reasoning for Context-Aware Reward Modeling of Robotic Endovascular Guidewire Navigation",
   "year": 2026,
   "venue": null
  },
  {
   "key": "liu2024_autonomous",
   "doi": null,
   "url": "https://arxiv.org/abs/2403.05748",
   "title": "Autonomous Guidewire Navigation for Robot-assisted Endovascular Interventions: A Knowledge-Driven Visual Guidance Approach",
   "year": 2024,
   "venue": null
  },
  {
   "key": "alekseeva2025_design",
   "doi": "10.1002/rob.22480",
   "url": null,
   "title": "The Design of Endovascular Robots for Cardiac Interventions: A Survey",
   "year": 2024,
   "venue": "Journal of Field Robotics"
  },
  {
   "key": "belikov2024_evolution",
   "doi": "10.1016/j.jestch.2024.101789",
   "url": null,
   "title": "Evolution and current state of robotic catheters for endovascular surgery: A comprehensive review",
   "year": 2024,
   "venue": "Engineering Science and Technology, an International Journal"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (song2022_biomimetic); partial (S8) for the rest",
 "notes": "NMPA registration not verified. Only commercial-grade robot with simulation-trained controllers on hardware (phantom only). | X2 2026-11-08: song2022 (Micromachines, CC BY 4.0) read in full: 18 DOF (12 was the short-axis count), 38 swine cases.",
 "maintainers_own": false,
 "found_by": "commercial"
}