 "schema_version": 1,
 "id": "aichi_miyachi2019",
 "name": "Aichi support robot",
 "full_name": "Neuroendovascular intervention support robot with insertion-force sensing (Aichi)",
 "group": "Aichi Medical University, Department of Neurosurgery and Neuroendovascular Therapy Center (Shigeru Miyachi) with Aichi University of Technology (Yoshitaka Nagano) and Nagoya City University (T. Hironaka) (miyachi2019_novel affiliations)",
 "group_short": "Aichi Med. U.",
 "family": "academic",
 "category": "academic",
 "first_year": 2019,
 "latest_year": 2025,
 "versions": "2019 prototype with 2 independent slaves and 2-joystick master (miyachi2019_novel, World Neurosurg 127:e617, full); 2021 remote operation over LAN (miyachi2021_remote, PMC HTML); 2025 review with miniaturised 3D-printed slave and automatic coil insertion (miyachi2025_robotics)",
 "target_anatomy": "cerebral (aneurysm coiling)",
 "devices": [
  "guidewire",
  "microcatheter",
  "catheter",
  "therapeutic-device"
 ],
 "devices_driven": "microcatheter (forward/back) and microguidewire (forward/back/twist) at once, by two independent slaves; guiding catheter placed manually (miyachi2019_novel Methods, Fig. 3); coil delivery in the 2025 version (miyachi2025_robotics Fig. 3)",
 "n_dof": 3,
 "n_dof_detail": "3 (MC translation; MGW translation + rotation) (miyachi2019_novel Slave Manipulator, Fig. 3)",
 "translation_mechanism": "friction-roller",
 "translation_detail": "friction-roller: device pinched between two rollers, DC motor through a 523:1 gear, upper roller in an open-close cover; roller grip set to slip above 1 N push (miyachi2019_novel Slave Manipulator, Driving Speed and Safety Function)",
 "rotation_mechanism": "other",
 "rotation_detail": "MGW roller unit and force sensor sit inside a drum turned by an engaging roller with a rubber anti-slip belt; battery inside the drum allows 360 deg rotation (miyachi2019_novel Slave Manipulator, Fig. 3); Bluetooth link to its motor in 2021 (miyachi2021_remote)",
 "device_range": "N.R.",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "proximal: push/pull force sensor behind the MGW roller actuator, shown on the monitor in newtons and as a musical tone; repeatability +/-0.04 N (miyachi2019_novel Slave Manipulator, Results, Figs. 4-5); insertion-force sensor in the Y-connector in later versions",
 "haptics_leader": "no haptics; two joysticks (left MC, right MGW; tilt left/right advances/withdraws, forward/back twists the MGW, speed proportional to tilt) with audio/visual force display (miyachi2019_novel Master Manipulator)",
 "imaging_compat": "X-ray angiography room: clinical simulation on the aneurysm model in a standard AngioSuite with the operator in the control room (miyachi2019_novel Fig. 7)",
 "sterility": "slave units are waterproof; electric cords and circuit boards under the supporting frame (miyachi2019_novel Slave design); drape N.R.",
 "highest_validation": "phantom",
 "validation_detail": "bench: > 50 master-slave trials without error motion, force-gauge calibration (miyachi2019_novel Results); phantom: microcatheterisation and aneurysm-sac insertion in a wet EVE silicone whole-vessel model with closed saline circulation under fluoroscopy, repeated AngioSuite trials all successful (Results, Figs. 6-7); remote coiling with the master 50 m away over LAN/HTTP (miyachi2021_remote)",
 "key_numbers": "slave output force > 1 N; guidewire speed about 1 mm/s (523:1 gear); slave roller slips above 1 N pushing load (safety); force gauge +/-0.04 N; start-motion lag 0.1-0.2 s (miyachi2019_novel); delays about 0.1 s each for motor stop, Bluetooth, video and network (miyachi2021_remote)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "B",
  "kind": "photo",
  "source": "miyachi2019_novel Fig. 2 (external appearance of the whole system), p. e618, World Neurosurg 2019 doi:10.1016/j.wneu.2019.03.221",
  "source_url": "https://doi.org/10.1016/j.wneu.2019.03.221",
  "licence": "publisher copyright (Elsevier)",
  "credit": "Miyachi et al., World Neurosurgery (2019), Fig. 2. publisher copyright (Elsevier).",
  "edits": "native extract (957 x 656 figure); cropped below the top label and above the joystick box; nothing erased; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "aichi_miyachi2019_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "photo",
   "caption": "Slave robot system: catheter robot, wire robot and insertion force sensor",
   "source_key": "miyachi2021_remote",
   "source": "Miyachi et al., Asian Journal of Neurosurgery (2021), Fig. 1 (PMC8244710 figure g001)",
   "source_url": "https://doi.org/10.4103/ajns.ajns_77_21",
   "where": "Fig. 1 (PMC8244710 figure g001)",
   "licence": "CC BY-NC-SA 4.0"
  },
  {
   "id": "aichi_miyachi2019_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Slave robot and vascular model on the table in the angiography room",
   "source_key": "miyachi2025_robotics",
   "source": "Miyachi et al., Journal of Neuroendovascular Therapy (2025), Fig. 5(B) (PMC11911030 figure g005)",
   "source_url": "https://doi.org/10.5797/jnet.ra.2024-0112",
   "where": "Fig. 5(B) (PMC11911030 figure g005)",
   "licence": "CC BY-NC 4.0"
  },
  {
   "id": "aichi_miyachi2019_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Slave side during remote coil embolization, angiography monitor behind",
   "source_key": "miyachi2025_robotics",
   "source": "Miyachi et al., Journal of Neuroendovascular Therapy (2025), Fig. 5(D) (PMC11911030 figure g005)",
   "source_url": "https://doi.org/10.5797/jnet.ra.2024-0112",
   "where": "Fig. 5(D) (PMC11911030 figure g005)",
   "licence": "CC BY-NC 4.0"
  },
  {
   "id": "aichi_miyachi2019_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Insertion force sensor head with curved through-hole and load cells",
   "source_key": "miyachi2021_remote",
   "source": "Miyachi et al., Asian Journal of Neurosurgery (2021), Fig. 2 (PMC8244710 figure g002)",
   "source_url": "https://doi.org/10.4103/ajns.ajns_77_21",
   "where": "Fig. 2 (PMC8244710 figure g002)",
   "licence": "CC BY-NC-SA 4.0"
  },
  {
   "id": "aichi_miyachi2019_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "Master side operator at the remote control room 50 m away",
   "source_key": "miyachi2025_robotics",
   "source": "Miyachi et al., Journal of Neuroendovascular Therapy (2025), Fig. 5(C) (PMC11911030 figure g005)",
   "source_url": "https://doi.org/10.5797/jnet.ra.2024-0112",
   "where": "Fig. 5(C) (PMC11911030 figure g005)",
   "licence": "CC BY-NC 4.0"
  },
  {
   "id": "aichi_miyachi2019_06",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Slave structure diagram and the inside of the MC and GW drivers",
   "source_key": "miyachi2019_novel",
   "source": "Miyachi et al., World Neurosurgery (2019), Fig. 3(A)-(B), p. e619",
   "source_url": "https://doi.org/10.1016/j.wneu.2019.03.221",
   "where": "Fig. 3(A)-(B), p. e619",
   "licence": "publisher copyright (Elsevier)"
  },
  {
   "id": "aichi_miyachi2019_07",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Force-sensor head on the coil delivery wire with the insertion-force readout",
   "source_key": "miyachi2019_novel",
   "source": "Miyachi et al., World Neurosurgery (2019), Fig. 4, p. e620",
   "source_url": "https://doi.org/10.1016/j.wneu.2019.03.221",
   "where": "Fig. 4, p. e620",
   "licence": "publisher copyright (Elsevier)"
  },
  {
   "id": "aichi_miyachi2019_08",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Slave beside the aneurysm model in the AngioSuite; operator at joysticks in the control room",
   "source_key": "miyachi2019_novel",
   "source": "Miyachi et al., World Neurosurgery (2019), Fig. 7, p. e622",
   "source_url": "https://doi.org/10.1016/j.wneu.2019.03.221",
   "where": "Fig. 7, p. e622",
   "licence": "publisher copyright (Elsevier)"
  }
 ],
 "sources": [
  {
   "key": "miyachi2019_novel",
   "doi": "10.1016/j.wneu.2019.03.221",
   "url": null,
   "title": "Novel Operation Support Robot with Sensory-Motor Feedback System for Neuroendovascular Intervention",
   "year": 2019,
   "venue": "World Neurosurgery"
  },
  {
   "key": "miyachi2021_remote",
   "doi": "10.4103/ajns.ajns_77_21",
   "url": null,
   "title": "Remote surgery using a neuroendovascular intervention support robot equipped with a sensing function: Experimental verification",
   "year": 2021,
   "venue": "Asian Journal of Neurosurgery"
  },
  {
   "key": "miyachi2025_robotics",
   "doi": "10.5797/jnet.ra.2024-0112",
   "url": null,
   "title": "Robotics for Neuroendovascular Therapy",
   "year": 2025,
   "venue": "Journal of Neuroendovascular Therapy"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (miyachi2019_novel, ); partial (miyachi2021_remote PMC HTML via WebFetch); full (miyachi2025_robotics)",
 "notes": "2026-11-08 X3: miyachi2019 read in full; main picture replaced by the same photo from the 2019 paper at 957 px with fewer labels (old 590 px file kept as aichi_miyachi2019_prev.png)",
 "maintainers_own": false,
 "found_by": "gap"
}